Power Switch -
mains disconnect/circuit breaker for the cabinet.
Blower Motor/Lights Reset Button
(15 Amp) - The Reset but-
ton (located on the left side of the control panel, directly above
the Receptacle fuses) is an in-line circuit breaker for the internal
blower motor and lighting. If an overload occurs, the circuit
breaker will trip and the button will protrude from the panel.
Depress the button to reset the circuit breaker.
Note:
• Turn power off to the blower.
• Turn power off to the lighting.
• Press the blower motor/lighting Reset button.
Receptacles
- Receptacles (115 Volts) are located on the left and
right sidewall of the workstation. Power is controlled by the
receptacle switch located on the control panel. The maximum
load is 5 amps total for Models 1284 and 1285. Models 1286
and 1287 have a maximum load of 2 amps total. Models 1285
and 1287 are equipped with 2 single European 230V receptacles.
Drain Valve
- The drain valve, located on the right front side of
the cabinet, is provided for the safe drainage of the drain pan.
This valve must remain closed while work is being performed in
the cabinet and be used only in the event of a major spill.
If a spill occurs, immediately consult a biological
safety officer or other qualified individual for
proper procedures. To contain a spill, connect a
sealed hose from the drain valve to a sealed con-
tainer.
Service Valves
- Two service valves are standard with each cabi-
net. These valves are located on the right and left side of the
workstation and can be coded with the type of service that they
supply. Identification index buttons are supplied.
The cabinet supports four service valves. Additional valves are
available.
Exhaust Filter Guard
- The exhaust filter guard, located on top
of the exhaust filter, protects the exhaust airflow and prevents
the storage of objects on top of the housing.
Sliding Window Assembly
- The sliding window assembly
allows the operator to raise the glass window to place items
within the work area. If the sliding window is above the 10-inch
level, a red light and audible alarm warns that an unsafe condi-
tion exists.
When work is being performed in the cabinet, the
sliding window must be at the 10-inch position to
avoid contamination to product and personnel.