13.2 Model 1286
Stock #
Description
156109
HP Blower Motor (1500 RPM)
170045
Capacitor, Motor 25MFD, 370V
190396
Motor Speed Control
500030
Ballast (Fluorescent Lighting)
141048
Fluorescent Lamp (32 Watt)
230054
Circuit Breaker, 15A SP
430263
Line cord Assembly, 20A, 120V,
Hospital Grade
760180
Filter, Supply HEPA 18” x 72” x
5-7/8” Pressure Range: 0.78”-1.05”
at 1350 cfm, filter drop pressure test
ed per IES-RP-CC001.3
760181
P/N P195118, Pressure Range:
0.30”-0.50” at 352 cfm, filter drop
pressure tested per IES-RP-CC001.3
500009 Ballast,
(UV Lighting)
141014
30W Germicidal Lamp
300305
Delay Relay, (fixed 5 min.)
280005
Pilot Light, #312, Red
360096
Push-button Switch, SPDT
360096
Rocker Switch, SPST, Flat Black
360105
Rocker Switch, SPDT
249025
Valve Body w/Tip
104008
Gauge, Static Pressure
360146
Rocker Switch, DPST
230166
Fuses, 3 Amp 1/4” x 1-1/4”
230204
Circuit Breaker/Switch 18A
The parts for the Model 1287 are the same as Model 1286
with the addition of:
Stock #
Description
275012
Transformer, 1.5KVA, 240/120V
430270
European Cord, 230V, 16A
230179
Circuit Breaker/Switch 10A
13.3 Filter Pressure Drop Conversion
Pressure drop across a HEPA filter is linear which allows
one to accurately predict the pressure drop at various CFM if
given a starting value. It is a straight proportion from one setting
to the other.
Example:
A filter rating of 0.31” of water at 352 CFM needs to be
converted to 530 CFM.
The formula is as follows:
0.31 = x
352
530
Solving for x (the needed pressure drop at 530 CFM)
x=(0.31)*530
352
x=0.47” of water (rounded)
13 -2
Model 1280 Series __________________________________________________________________________Parts List