28
Art. no.: A005-962_Rev 01
Last update: CW34/2016
THERA-Trainer coro - User manual
OPERATION
The tread extension can consist of the following parts:
Connection part, magnetic
1
Mid part, magnetic
2
Ramp, magnetic
3
Connect the individual components in the right order.
Connect the connction part
1
with the tread unit.
Connect the mid part
2
with the connection part
1.
Connect the ramp
3
with the mid part
2.
To install, place segments together so that they are flush and connect them
magnetically.
Only transport THERA-Trainer coro without tread extension.
8.10. Setting balance function
Set release lever
1
to position 0, 1 or 2 on both vertical pipes.
–
Position 0
= balance function disabled
4
.
–
Position 1
= balance function enabled with limited range of movement
(up to 6°)
3
.
–
Position 2
= balance function enabled with full range of
movement (up to 11°)
2
.
Ensure that release lever
1
engages.
Set balance function to the same position on both sides.
WARNING!
Risk of injury due to insufficient preparation!
Before every training, ensure that the balance unit is functioning.
Lift THERA-Trainer coro left and right onto the upper pipe frame.
If you can lift the vertical pipe out of the balance unit, then do not train.
Contact trained suppliers.