background image

Texmate, Inc. Tel. (760) 598-9899  •  www.texmate.com

FL-B101Q-DCA _ FL-B01Q-DCV (d0084)

Page 9

To calibrate the Analog Output you must be able to input two input signals. Usually the mini

-

mum input (LO Input) and the maximum (HI Input) signals are used for maximum 

accura

-

cy.

For example the five steps to obtain an Analog Output of 4mA to 20mA  for an input of 0 to 

10V are:

STEP A  ACCESS THE ANALOg CALIBRATION MODE

 

1)  Confirm the internal analog output module is installed and that the required 

voltage or current output option is selected.

 

2)  Turn OFF the power to the bargraph.

 

3)  Hold down the  ZERO and SPAN buttons simultaneously and re-power the 

 

bargraph. The ZERO button will now function as the LO button and the SPAN 

button will now function as the HI button for calibrating the Analog Output.

STEP B  RESET THE ANALOg OuTPuT SCALINg

 

1)  Press  the  LO  and  HI  buttons  simultaneously  and  hold  them  down  for  2 

 

seconds. This will reset the analog output scaling to the default value. The 

default analog output scaling is approximately 0 to 20mA (0 to 10V if voltage 

output option is selected) for an input that is 0 to 100% of the range selected 

on the input signal conditioner.

STEP C  CALIBRATE  ANALOg OuTPuT FOR LO SIgNAL

 

1)  Apply the low input signal (0V in this example) to the meter.

 

2)  Connect an external multimeter to the analog output pins (Pins 17 and 18).

 

3)  Using the LO button adjust the analog output as measured on the external 

 

multimeter  to  be  the  required  value.  (4mA  in  this  example).  When  the  LO 

 

button  is  pressed,  the  UP  or  DOWN  indicator  LED  shows  the  direction  of 

change.  To reverse the direction of change release the LO button and press 

down again.  Initially the output changes very slowly, but speeds up as the LO 

button remains pressed down. The analog output for a low input can be set 

in this step to any value in the range of  0 to 20mA or 0 to 10V ( if the voltage 

output option is selected).

STEP D  CALIBRATE ANALOg OuTPuT FOR HI SIgNAL

 

1)  Next apply the high input signal (10V in this example) to the meter.

 

2)  Using the HI button, adjust the analog output as measured on the external 

 

multimeter  to  be  the  required  value.  (20mA  in  this  example).  When  the  HI 

 

button  is  pressed  the  UP  or  DOWN  indicator  LED  shows  the  direction  of 

change.  Release the HI button and press again to reverse the direction of 

change. Initially the output changes very slowly, but speeds up as the HI button 

continues to remain pressed. This output may be higher or lower than the value 

set in Step C, and may be any value in the range of 0 to 20mA or 0 to10V. This 

allows the easy reversal of analog output that is required in some applications.

STEP E  EXIT THE ANALOg OuTPuT CALIBRATION MODE

 

1)  Turn OFF the power to the bargraph

 

2)  Re-power  the  bargraph.  The  two  buttons  will  now  return  to  their  original 

 

function of ZERO and SPAN. 

 

3)  Calibration  is  now  complete  and  the  bar  is  scaled  for  a  0  to  10V  input  to 

 

produce an analog output of 4 to 20mA.

Turn Power ON 

while holding 

down the Zero

and Span Buttons

simultaneously

100

0

SP1 SP2 SP3 SP4

Zero Span

Apply 0 V to the 

Input Signal Pins

100

0

SP1 SP2 SP3 SP4

Zero Span

+

Adjust the Analog

output to 4.00mA with

the LO (zero) button

17

18

4.00

Apply 10 V to the 

Input Signal Pins

100

0

SP1 SP2 SP3 SP4

Zero Span

Adjust the Analog 

output to 20.00mA with

the HI (span) button

17

18

20.00

+

Turn Power OFF 

and then back 

ON to exit 

Analog output 

Calibration Mode 

100

0

SP1 SP2 SP3 SP4

Zero Span

When the optional analog output module is installed, an independently calibrated 16 bit isolated, voltage or current analog output is available. 

The analog signal is independently scaled to the input signal and not to the bargraph display. It is important to note that the Analog 

Output is completely independently of the bargraph display.  This means for example that the bargraph display may be scaled to go from zero 

to full scale as the input changes from 0 to 5V, while at the same time, the analog output is scaled to go from 4 to 20mA as the input changes 

from 2 to 3V. Rescaling the bargraph or the analog output will not affect the scaling of the other.

Analog Output Scaling and Calibration

Reset the analog

output scaling by

pressing the LO (zero)

and HI (span) buttons

simultaneously

for 2 secs.

100

0

SP1 SP2 SP3 SP4

Zero Span

Summary of Contents for FL-B101Q-DCA

Page 1: ...s Analog Output Isolated 16 bit user scalable mA or V OIC mA out 4 20 mA 0 to 500 max loop resistance OIV volts out 0 10 V DC 500 or higher resistance Power Supply AC DC Auto sensing wide range supply...

Page 2: ...As there are no menu or sub menus to navigate the programming and setup is quick and Span Buttons Span Zero Buttons Zero Set Point Button SP1 SP2 SP3 SP4 Front Panel Buttons Zero Button The Zero Butt...

Page 3: ...alled Mode 3 Hysteresis Band between SP1 SP2 This mode enables the Hysteresis function In order for Hysteresis to function SP2 must be set to a value greater than SP1 and SP2 should be selected as Hig...

Page 4: ...ver and carefully lever the plastic catch up and tilt the Display Board assembly forward Then repeat this action with the bottom catches Step 5 Remove the entire Display Board assembly by sliding it o...

Page 5: ...or SPAN but ton is pressed the input signal being applied must be more than 5 higher or lower than the previously memorized value of the input signal that was being applied when the other button was...

Page 6: ...to 20mA If now the scaling has to be changed to read zero to full scale for an input of 0 to 20mA only the low 4 mA end needs to be recalibrated The high 20 mA end of the scaling is left untouched an...

Page 7: ...2 and SP4 activate when input is lower than set point SP1 and SP3 activate when input is equal to or higher than set point SP2 activate when input is lower than set point SP1 SP3 and SP4 activate when...

Page 8: ...SP2 color overrides the SP1 color STEP A ENTER COLOR SET MODE Hold down ALL four setpoint buttons SP1 SP2 SP3 and SP4 and apply power to the meter The meter will light up in the Color Set Mode Release...

Page 9: ...HI button adjust the analog output as measured on the external multimeter to be the required value 20mA in this example When the HI button is pressed the UP or DOWN indicator LED shows the direction o...

Page 10: ...T mA 0 to 20 mA 4 to 20 mA or V 0 to 10 V output is header selectable Pin 19 Programming LOCK By connecting the LOCK pin to the COMMON pin the meter s programmed parameters can be viewed but not chang...

Page 11: ...Pot Span Pot ID05 ID01 Zero Offset Pot Zero Offset Pot Input Range Header Input Range Header Input Range Header ID01 DC Volts 2 20 200V Custom w 24V DC Exc Custom 200V 20V 2V ON OFF 24V Exc 24V Exc D...

Page 12: ...or FI series IO board The adapter snaps on the 36x144 mm 1 42 x5 69 case and enables single unit or stack mounting in an existing 6 edgewise pointer meter cut out 100 0 25 50 75 100 0 25 50 75 SP1 SP2...

Page 13: ...into a Hi and a Lo range This has the effect of dividing the adjustment range of the SPAN pot into ten equal 10 steps across 100 of the input Signal Span Range HI LO HI LO SPAN Turn Clockwise to Incre...

Page 14: ...5 150 1500 2 20 200 2000 2 5 25 250 2500 3 30 300 3000 4 40 400 4000 4 5 45 450 4500 5 50 500 5000 6 60 600 6000 7 5 75 750 7500 8 80 800 8000 9 90 900 9000 1 2 12 120 1200 Div Face plate scales and n...

Page 15: ...g 5 93 PLUG5P DR Extra Screw Terminal Conn 5 Pin Plug 6 DN CAS144X36 Complete 144x36mm Case with bezel 25 OP MTL144X36 Metal Surround Case includes screw mounting clips 25 OP MTLCLIP Screw Mounting Cl...

Reviews: