The DL-40JANUS is an intelligent meter with a hierarchical
software structure designed for easy programming and opera-
tion, as shown below in the software logic tree.
After the meter has been powered up, the four
digits light up for three seconds and then settle to
the operational display indicating the input signal.
15 Second
Program Timeout
The meter has a 15 second
program timeout. If no buttons
are pressed for 15 seconds, at
any stage of the programming
sequence the meter will exit the
programming mode and return
to the operational display. Any
program changes that were
made prior to pressing the
P
button in the preceding step
will not be saved.
Software Logic Tree
SETPOINT SETTING AND
RELAY CONFIGURATION MODE
See Page 9
Set Setpoint 1
(SP1)
Delay-on-Make
(doM)
Delay-on-Break
(dob)
Setpoint 2
(SP2)
Hysteresis
(HYSt)
Hysteresis
(HYSt)
Hysteresis
(HYSt)
Hysteresis
(HYSt)
Delay-on-Make
(doM)
Delay-on-Break
(dob)
Setpoint 3
(SP3)
NOTE: [dob] [dom] Functions
are only available
for SP1 and SP2
Setpoint 4
(SP4)
Relays Activation [rLYS]
(H) High the relay energizes
when the setpoint is exceeded.
(L) Low the relay energizes below
the setpoint. Setpoint are indicated
from left to right SP1, SP2, SP3, SP4
Peak
Reset
PEAK
Reset
VALY
Setpoint 1
(SP1)
confirm (SP1)
confirm
Peak
confirm
Valley
confirm Analog
output values
confirm (SP2)
confirm (SP3)
Setpoint 2
(SP2)
[LHLH]
[HLHL]
[HHHH]
MAIN MENU
Operational Display
SETPOINT
VIEW ONLY MODE
PEAK & VALLEY
VIEW & RESET
Sub-menu
MODE
Calibration
Mode
Calibration
Mode
DECIMAL POINT AND
BRIGHTNESS SELECTION
See Page 6
Setpoint 3
(SP3)
Setpoint 4
(SP4)
Valley
Offset
Scale
Factor
Calibrate
Analog
Output
Lo
Calibrate
Analog
Output
Hi
[X•XXX]
[XX•XX]
[XXX•X]
[XXXX•]
[XXXX]
[2]
[3]
[4]
Decimal Point
(dp)
Display
Brightness (br)
TWO POINT ANALOG OUTPUT
CALIBRATION
SEE PAGE
6
MANUAL
RESCALING MODE
See Page 4
SELECTION FOR ANALOG
RANGE OUTPUT
See Page 6
confirm (SP4)
confirm
confirm
confirm
This menu only available
when an analog output
module is installed
When MANUAL
is set to ON
When AUTO
is set to OFF
Span
Zero
Err. Any new setting
canceled and previous
settings are retained
TWO POINT AUTO
CALIBRATION MODE
SEE PAGE 5
This menu only available
when an analog output
module is installed
Modbus RS-485 Settings
See Page 7-8
BAUD: -press Up button for 192(00)
-press Down button for 96(00)
ADDR: -press Up or Down button to change
from 1 to 247.
-value rolls over, e.g. 247 to 1
PRTY: -press Up or Down button to change
-values are 2=Even, 1=Odd, 0=None
Texmate’s Modbus RTU RS-485 module provides
half-duplex, serial communication to PLC, SCADA and
other systems for monitoring and control of the meter’s
functions. The meter functions as a slave.
The RS485 module takes the place of the Analog Output
module, so it is not possible to have both Analog
Output and RS-485 communication. Programming and
operation of the meter are done via a different menu
system, or Software Logic Tree, that is accessed through
the front panel push buttons, see below.
14 15
8
9
10 11
See Leopard Family Input
Signal Conditioning Modules
1
2
3
4
5
6
Relay Outputs
AC
Neutral
– DC
AC
Line
+ DC
(Lock) ON
COM Ref A+
29 28 27 26 25 24 23 22
B
1
2
With RS-485 Output Option
MANUAL rescaling- (
without external signal
) ON
AUTO
scaling
- (
with external signal
) OFF
17 16
19 18
ON
SP1
NO
SP1
NO
SP2
NO
SP2
NO