10
7 CC indicator
Indicates during constant-current operation.
8 OUTPUT indicator
Red LED indicates when the output is on. The preset voltage is available at the output
terminals when this indicator is on.
9 OUTPUT switch
Output switch (contactless) that turns on and off the output electrically.
When the output is on, the OUTPUT indicator 8 lights and the voltage the voltmeter is
reading is output at the output terminals and the ammeter reads the current.
10 V/I CHECK
Voltage/current check switch. As long as this switch is depressed, the voltmeter reads
the preset voltage and the ammeter the preset current so that you may preset voltage
and current. To check the preset voltage and current, depress this switch when the
OUTPUT switch is on.
11 OVP CHECK
Overvoltage protection level check switch. Depress this switch and the voltmeter
reads the preset overvoltage protection level so that you may preset the level with the
OVP adjuster
⑫
.
12 OVP adj. control
Semi-fixed adjustment control permitting to preset the overvoltage protection level.
13 Output terminal (+)
Positive output terminal (red).
14 Output terminal (-)
Negative output terminal (white).
15 GND
Ground terminal connected to the frame. This is connected also to the output terminal
(-) normally.