Test Data and Simulation Results
www.ti.com
6.3
FFT Results
Figure 32
and
Figure 33
show the FFT results of the ADS1293 device corresponding to different peak
rates.
Figure 32. FFT Plot ECG Channel
Figure 33. FFT Plot Pace Channel
(50-Hz Signal)
(50-Hz Signal)
6.4
EMI Test Results of the ADS1293
Table 2. Test Results
DUT
Frequency MHz
Vos_Off (uV)
Vos_On (uV)
Vrf_pp (mV)
EMIRR
4L2
400
-4.94
-4.93
130.8
132.6231099
4L2
900
-4.82
-5
103.4
103.4341716
4L2
1800
-5.05
-4.98
90
109.2265398
4L2
2400
-4.95
-4.9
45.2
100.1849375
6L1
400
19.03
18.99
218.1
129.463827
6L1
900
19
18.97
225.6
132.5499389
6L1
1800
19
18.98
185.7
132.6906764
6L1
2400
19.01
18.99
41.9
106.8267612
11L1
400
-4.55
-4.36
204.1
114.7774483
11L1
900
-4.38
-4.26
204.8
118.8283733
11L1
1800
-4.39
-4.48
147.2
115.5902624
11L1
2400
-4.42
-4.37
47.73
101.1310575
24
Wireless Heart Rate Monitor Reference Design
TIDU195A – January 2014 – Revised July 2014
Submit Documentation Feedback
Copyright © 2014, Texas Instruments Incorporated