Testing the PGA5807A EVM
4. Return to the
PGA5807
tab and change the
PGA_GAIN
in the
SPI CONTROL SECTION
to
0 dB
as
shown in
.
Figure 46. PGA5807 GUI Setup for PG ADS5296A SNR, THD, Gain Test 2
5. Clicking on the
ADC
tab and pressing
Capture
results in
. Take the difference in the
Fund.
value shown here to the previous capture. This difference should equal the difference in
PGA_GAIN
values between the two captures, or 18 dB. In this example, the difference is 17.4 dB.
34
PGA5807A, 8-Channel, High-Bandwidth, Analog Front-End Evaluation
SLAU538 – October 2013
Module
Copyright © 2013, Texas Instruments Incorporated