Other Items Needed for Operation
14
DLPU016C – April 2014 – Revised April 2019
Copyright © 2014–2019, Texas Instruments Incorporated
DLP NIRscan Evaluation Module Overview
Table 1-2. DLP NIRscan EVM Components
Electronics
Device Number
Description
DLP chipset
DLPC350
Controller for the DLP4500 family of DMDs
DLPC4500NIR
DMD
Embedded processor
AM3358
Sitara processor
Analog chain for detector
ADS1255
24-bit, 30 kSPS very-low noise delta-sigma ADC
OPA2376
Precision, low noise, low quiescent current operation amplifier
Analog chain for TEC driver
and lamp driver
OPA569
Power operational amplifier, output signal swings within 200 mV of rails at 2-A output
INA330
Thermistor signal amplifier for temperature control
OPA340
Micro operational amplifier for PID control of TEC driver
Power management
TPS65217C
Single-chip PMIC for battery-powered systems for AM3358, DDR3
TPS65145
4-channel LCD bias with fully-integrated positive charge pump, 3.3-V LDO controller for
DLP4500NIR power
TPS73025
Single output LDO, 200 mA, fixed (2.5 V), high PSRR, low noise for DLP4500NIR power
TPS65251
PMU with three DC-to-DC converters, up to 18-Vin for DLPC350 power
TPS79718
Single output LDO, 50 mA, fixed (1.8 V), low quiescent current, Powergood out for
DLPC350 power
TLV61230
5-V supply for USB and input to TPS65217
TLV61230
3.3-V supply
REF5025
Low noise, very-low drift, precision voltage reference for TEC and ADC references
TPS71750
5-V analog supply for ADC and TEC
TPS71733
3.3-V analog supply for ADC
ESD protection
TPD4S012
4-channel USB ESD solution with power clamp
1.3
Other Items Needed for Operation
The DLP NIRscan EVM spectrometer is a flexible, ready-to-use EVM. However, the DLP NIRscan EVM
does not ship with cables, power supply, or additional hardware components. To use the EVM, the user
needs:
•
Power supply:
–
Nominal voltage: 12-V DC
–
Typical current: 3 A
–
Maximum current: 5 A
–
DC connector size:
•
Inner diameter: 2.5 mm
•
Outer diameter: 5.5 mm
•
Shaft: 9.5-mm female, center positive
–
Efficiency level: V
–
A recommended power supply is Digi-Key
®
, or equivalent
•
USB cable: A-to-mini-B USB cable
•
External laptop, desktop, or personal computer
1.4
DLP NIRscan Connections
and
depict the connectors with their respective locations. No cables, nor power
supply are included with the unit. Only connection 8 and connection 14 or 7 are required for typical
operation. All of the connections are listed for completeness in