Upgrading the DLPC350 Firmware
39
DLPU016C – April 2014 – Revised April 2019
Copyright © 2014–2019, Texas Instruments Incorporated
NIRscan EVM Reprogramming Procedure
Figure C-3. LED Location
6. Reprogramming typically occurs in fewer than 10 minutes. When programming completes successfully,
D1 and D2 stop flashing and stay illuminated.
7. After confirming that D1 and D2 have illuminated steadily for at least 20 seconds, remove power from
the board by removing the power cable.
8. Remove the microSD card from the board.
9. Reapply power to the board by plugging in the cable.
10. The NIRscan boots from the new software. For the first operation after reprogramming, the NIRscan
boot takes slightly longer to complete. When LED D1 illuminates, it indicates that the NIRscan is ready
to communicate. Plug in the USB cable, enter
http://192.168.0.10
into the browser URL window, hit
enter, and the NIRscan menu appears.
C.3
Upgrading the DLPC350 Firmware
This section describes how to reprogram the DLP NIRscan EVM with its application software. Because the
NIRscan firmware version and the DLPC350 version have dependencies, when upgrading the DLP
NIRscan EVM software, the DLPC350 may also require an upgrade.
These are the two compatible versions:
•
NIRscan version 1.0.0 operates with DLPC350 firmware version 1.1 only.
•
NIRscan version 2.0.0 operates with DLPC350 firmware version 2.0 only
The steps to reprogram the DLPC350 are:
1. Download the latest
select save rather than open to a directory and
remember the file location. For example, if reprogramming the NIRscan with Application software
version 2.0.0, select DLPR350PROM version 2.0