2
Reference
ADS126xEVM Software
3.4.2
Tab 2: Reference
shows the tab used to configure the ADCx reference source. The voltage reference source is
selected from the
ADCx REF Source
drop-down menu. When using an external source for ADC1, the
positive and negative reference inputs must be specified by the
REFP
and
REFN
drop-down menus.
Selecting the
Invert ADC1 REF Inputs
checkbox swaps the positive and negative reference inputs, and
allows for fully flexible reference source inputs.
Figure 18. Tab 2 Settings
When using an external reference source for ADC1, also make sure that the
ADC1 REF Voltage
field is
matched to the applied reference voltage to allow ADCPro to correctly convert the output data from codes
to volts. The internal reference can be disabled when an external referenced is used; however, the internal
reference must be enabled to use the IDACs.
The
ADC2 Reference Settings
are only visible when an ADS1263EVM is connected.