Adjusting parameters in the device
25
OpenOffice/LibreOffice Calc is used, this row appears after the im-
port. It can be deleted. After that comes the prefix, so that the column
titles are named in Office programs. This is followed by the actual
data.
The file format for messages is similar:
sep=,
"error message","date","time",
AL,25 Change pump head 1,24.06.2020,10:26
AL,24 Indicator low,24.06.2020,10:26
Connection between time and correct data
For the names of the files and the date and time information in the file
to be correct, the clock must be set and working. If the battery is de-
pleted or no time is set, the date 1.1.2011, 12:00 is assumed auto-
matically and the data is saved. Data is not lost, as new measured
values and errors are appended to the existing files. However, only
one file is then written to at a time, since month and day changes do
not occur.
Functions of the USB connection
Installing the USB driver
Install the appropriate USB driver on the computer so you can use
the USB interface. You can find the VCP driver suitable for the oper-
ating system on the
USB connection to several devices
If the same computer is connected to other devices, the driver de-
fines a new COM port for each device!
Establishing a connection between the Testomat
®
Modul CL and
the computer
Connect the computer and the Testomat
®
Modul CL with a USB
cable.
program on the computer (see page 27).
Start selection mode on the Testomat
®
Modul CL with both func-
tion keys
Manual
and
Alarm
on page 22).
Start the USB communication (see
on page 22).
The yellow LEDs above the
Manual
and
Alarm
function keys light up
when connecting.
Now data can be displayed and processed by the device in the Ser-
vice Monitor program.
NOTE
NOTE