background image

8 Carrying out a test 

10

 

Attention: Use of the probe tip protector may be required depending on the 
national regulations or provisions! 

Probe tip protector: push onto probe tips or pull off. 

Probe tip extension: screw onto probe tips or unscrew. 

 

8.2.

 

Voltage testing 

 

 

Connect both probe tips to the test object. 

The instrument switches on automatically at a voltage of approximately 
6 V or above. 

The voltage is shown in the LC display. 

In the case of DC voltage, the polarity of the indicated voltage relates 
to the voltage tester probe tip. 

When the safety extra-low voltage is reached or exceeded (50 V AC / 
120 V DC), an acoustic signal is emitted, the red LED comes on and 

 is illuminated in the LC display. 

 

8.3.

 

Single pole phase testing (only testo 755-

2) 

Single pole phase testing is possible as from AC voltages of approx. 90 V. 
During single pole phase testing to determine external conductors, the 
display function may be impaired, for example due to insulating personal 
protective equipment or other insulators. 
Single pole phase testing is not suitable for testing for absence of voltage. 
Two-pole voltage testing is required for this. 

Connect a probe tip of the voltage tester to the test object.  

 is illuminated to signify phase testing on the relevant conductor. 

 

8.4.

 

Current measurement 

 

 

 

Strong interferences in the vicinity result in an unstable display 
and measurement errors. 

 

 

 

 

 

No voltage must be applied to the probe tips, so that the instrument 

switches to current measurement mode. 

Push the instrument's fork over the live conductor as far as the sensor 
zone. 

The reading is shown in the LC display. 

 

8.5.

 

Continuity/resistance testing 

 

Disconnect the test circuit/object from the power supply.  

 

Conduct a two-pole voltage test on the test object to confirm the 

absence of voltage. 

Connect both probe tips to the test object. 

For continuity up to approx. 30 

Ω, an acoustic signal is emitted, for 

resistance up to approx. 100 

kΩ the acoustic alarm remains inactive. 

The instrument switches off automatically after 10 seconds if no 
continuity/resistance is detected. As soon as continuity/resistance is 
detected, the instrument switches back on automatically. 

Summary of Contents for 755-1

Page 1: ...testo 755 Current voltage tester Instruction manual ...

Page 2: ...ng measuring point illumination on off 9 7 3 Switching the instrument off 9 8 Carrying out a test 9 8 1 Preparing the test 9 8 2 Voltage testing 10 8 3 Single pole phase testing only testo 755 2 10 8 4 Current measurement 10 8 5 Continuity resistance testing 10 8 6 Rotating magnetic field direction detection only testo 755 2 11 9 Service and maintenance 11 9 1 Replacing the batteries 11 9 2 Mainte...

Page 3: ...ranteed The fork meter must not be used while its battery compartment is open Batteries must be checked before use and changed if necessary If there is any battery leakage the instrument must no longer be used until it has been checked by our Customer Service The battery acid electrolyte is highly alkaline and electrically conductive Risk of acid burn If the battery acid comes into contact with yo...

Page 4: ...tage type directly or indirectly 4 Intended use The instrument may only be used under the conditions and for the purpose for which it was designed Current measurement voltage testing in the AC DC range of 6 to 600 V testo 755 1 or 6 to 1000 V testo 755 2 continuity testing resistance testing Only use the instrument within the specified measuring ranges of the following overvoltage categories Volta...

Page 5: ...l load Approx 3 5 W at 1000 V Current Is 3 5 mA at 1000 V Operating time 30 s Recovery time 240 s Automatic power on 6 V Reading memory HOLD testo 755 1 6 600 V AC DC testo 755 2 6 1000 V AC DC Overload indicator testo 755 1 630 V AC DC LC display shows OL testo 755 2 1050 V AC DC LC display shows OL Measurement category CAT III 1000 V CAT IV 600 V Single pole phase testing only testo 755 2 Featur...

Page 6: ...30 Ω to 100 kΩ Tolerances 1 of the display value 5 digits Test current 5 μA Overvoltage protection 1000 V AC DC Automatic power on 100 kΩ General technical data Feature Values Operating ambient temperature 10 C to 50 C Storage ambient temperature 15 C to 60 C Humidity Max 75 RH Operating altitude Up to 2000 m Level of contamination 2 Protection class IP 64 Power supply 2 x 1 5 V AAA IEC LR03 Power...

Page 7: ...w voltage exceeded Single pole phase testing 5 LC display Display Meaning AC AC voltage is applied DC DC voltage is applied HOLD Reading is recorded V Voltage in V A Current in A Safety extra low voltage exceeded 50 V AC 120 V DC Single pole phase testing only testo 755 2 phase detected Ω kΩ Resistance in ohms or kilo ohms Continuity Rotating magnetic field direction left or right ...

Page 8: ...sockets for probe tips 6 2 Explanation of icons Icon Meaning Attention Warning about a danger spot refer to instruction manual Caution Dangerous voltage risk of electric shock Continuous double or reinforced insulation in accordance with category II DIN EN 61140 Suitable for work on live parts Conformity mark verifies compliance with the valid EU Directives EMC Directive 2014 30 EU with the EN 613...

Page 9: ...hat the instrument is functioning perfectly for example at a known voltage source before and after every test If the safety of the user cannot be guaranteed the instrument must be switched off and secured to prevent unintentional usage Carrying out a function test Press HOLD for approx 2 s Instrument carries out a self test All segments of the LC display and the alarm along with the measuring poin...

Page 10: ...tive equipment or other insulators Single pole phase testing is not suitable for testing for absence of voltage Two pole voltage testing is required for this Connect a probe tip of the voltage tester to the test object is illuminated to signify phase testing on the relevant conductor 8 4 Current measurement Strong interferences in the vicinity result in an unstable display and measurement errors N...

Page 11: ...LC display 1 Fully disconnect the instrument from the measurement object 2 Using a screwdriver unscrew the two metal screws on the battery compartment until the battery compartment cover can be removed Do not unscrew the screws completely 3 Remove the spent batteries 4 Insert new batteries type AAA IEC LR03 1 5 V ensuring correct polarity 5 Put the battery compartment cover back on and screw down ...

Page 12: ... used until it has completely dried 10 Protecting the environment Dispose of faulty rechargeable batteries spent batteries in accordance with the valid legal specifications At the end of its useful life send the product to the separate collection for electric and electronic devices observe local regulations or return the product to Testo for disposal ...

Page 13: ......

Page 14: ...0970 7550 en 04 ...

Reviews: