Protection cap, measurement chamber, cooling coil
If used in process conditions involving oil or dust, the protection cap
and, if used, the measurement chamber and cooling coil must be
cleaned regularly.
•
Unscrew/remove protection cap, measurement chamber and
cooling coil and blow out with compressed air or place in an
ultrasonic bath.
•
Replace a heavily soiled protection cap with the same type of
protection cap.
The accuracy is only maintained when the ordered
protection cap is replaced by the same type of protection
cap.
Sensor cleaning
During sensor cleaning, avoid any contact with the sensor.
Do not mechanically clean the sensor, as otherwise the
cover electrode may be damaged.
1.
Carefully unscrew protection cap.
2.
Carefully rinse with isopropanol and/or distilled water.
3.
Allow sensor to dry thoroughly.
4.
Screw on protection cap.