4 Transmitter
39
8.
Press
SET
. Read off multimeter display (e.g. 18,401 mA) and
enter this value in the user menu. Do this by scrolling one digit
to the right using
and increasing the value of digit by 1 using
. Confirm with
SET
or abort entry with
ESC
.
9.
Return to
Main Menu Adjust
with
ESC
.
10.
Continue to
Main Menu Reset
with
or
or return to
Measuring Mode with
ESC
.
Pos: 55 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Reset bearbeiten 635x @ 3\mod_1238480830340_79.doc @ 30373 @ 4 @ 1
4.4.5.8.
Editing Reset main menu
You can reset the factory settings for the following:
• Instrument
•
Min./max. values
Resetting to the factory settings means resetting to the
order specification, i.e. the specific condition at the time of
supply to the customer.
1.
In the Measuring Mode press
SET
, select
Main Menu
Reset
with
or
and confirm selection with
SET
.
-
Reset Device
to factory settings is displayed.
2.
Select the setting to be reset using
or
and confirm
selection with
SET
.
-
Reset Completed
is displayed.
3.
Return to the reset setting with
ESC
or
SET
and return to
Main
Menu Reset
with
ESC
.
4.
Continue to the
Main Menu
Channel 1
with
or
or return to
Measuring Mode with
ESC
.
Pos: 56 /TD/Überschriften/MUF/4.5 Status-, Warn- und Fehlermeldungen @ 3\mod_1234442925442_79.doc @ 24963 @ 2 @ 1
4.5.
Status, warning and error messages
Pos: 57 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Status-, Warn- und Fehlermeldungen _6781 @ 3\mod_1241441355544_79.doc @ 32477 @ @ 1
To achieve optimum operational reliability (machine availability),
the transmitter shows the following via the user menu or the P2A
software
• Status
messages
• Warning
messages
• Error
messages
in each case for the testo 6781 transmitter.
All messages are stored in the transmitter with an operating hours
stamp. Use the user menu (see Editing Message main menu page
35) or the P2A software (see Transmitter history page 69) to view
the message history.