testo 6681
- 1.5 Status, warning and error messages
65
1.5.5 Namur fault conditions
If the faults named in the following table occur, the analog outputs output
special values that enable a general fault warning in the higher-level control
system. The values correspond to the "Namur" industry standard.
Analog output
Display message
Class
0-20mA 4-20mA 1V
5V
10V
No probe signal
Error
21mA 21mA
1.2V
5.5V
11V
Wrong probe
Error
21mA 21mA
1.2V
5.5V
11V
Watchdog error
Error
21mA 21mA
1.2V
5.5V
11V
Values < 0 % RH
Underrange
0mA 3.8mA
0V
0V
0V
Condensation
Overrange 20.5mA 20.5mA 1.2V
5.5V
11V
%RH short-circuit
Error
21mA 21mA
1.2V
5.5V
11V
%RH sensor broken
Error
21mA 21mA
1.2V
5.5V
11V
T short-circuit
Error
21mA 21mA
1.2V
5.5V
11V
T sensor broken
Error
21mA 21mA
1.2V
5.5V
11V
Probe disconnected
Error
21mA 21mA
1.2V
5.5V
11V