testo 6681
- 1.4 Operation
Enter reference value for 1-point adjustment
Please also refer to
chapter 1.3.3.2
1
In Measuring Mode, press SET, select
Main Menu Adjust
using
or
and confirm selection with SET.
Reference value % RH
is displayed.
2
Press SET, edit value: Scroll one digit to the right using
, increase
value of digit by 1 using
. Confirm with SET or cancel entry with
ESC.
3
Continue to
Reference value temp
with
or
.
4
Press SET and
Reference value °C
is displayed.
5
Press SET, edit value: Scroll one digit to the right using
, increase
value of digit by 1 using
. Confirm with SET or abort entry with ESC.
6
Continue to
Reference value °F
with
or
.
7
Press SET, edit value: Scroll one digit to the right using
, increase
value of digit by 1 using
. Confirm with SET or abort entry with ESC.
8
Continue to
Analog Adj. Ch. 1
with
or
.
9
Continue with the adjustment of the analog outputs (see below, step 2)
or press ESC to return to
Main Menu Adjustment
.
10
Continue to
Main Menu Reset
using
or
or return to Measuring
Mode by pressing ESC.
Performing the 2-Points-Adjustment for 20% RH and 80 % RH
1.
In the Measuring Mode press SET, select main menu Adjust with
or
and confirm selection with SET.
2.
Continue to 2Pt Adjust 20/80% with
or
.
3.
Press SET.
4.
Continue to Adj. Point 20% or Adj. Point 80% with
or
5.
Confirm selection with SET.
The red LED ADJ flashes
In the display
2Pt Adjust 20/80%
,
1-point adjustment
,
Probe reset
appears consecutively.
The adjusted measurement values appear after probe reset.