![TESTO 6681 Profibus Instruction Manual Download Page 136](http://html1.mh-extra.com/html/testo/6681-profibus/6681-profibus_instruction-manual_1092404136.webp)
136
P2A software
-
3.3 Using the software
Field
Explanation
Current
reading
Reading in % RH.
Readings are updated every second.
Reference
value
Entry of the read-off value from the reference measuring
instrument.
Permissible entries:
-
Lower adjustment point 10.3 to 12.3 % RH
-
Upper adjustment point 74.3 to 76.3 % RH.
6
Click on
[OK]
or
[Cancel]
to close the dialogue.