4 Transmitter
73
Pos: 74 /TD/Produkt verwenden/MUF 63xx/Hauptmenü Analyse bearbeiten 638x @ 3\mod_1234525515962_79.doc @ 25379 @ 45555
4.4.6.6.
Editing Main Menu Analysis
You can test the functionality of analog and relay outputs. In
addition, you can read off the minimum and maximum values (since
the last voltage supply or reset of the min./max. values).
Testing functionality of analog outputs
This function affects the analog outputs directly, not only
the test contacts.
1.
In the Measuring Mode press
SET
, select main menu
Analysis
with
X
or
S
and confirm selection with
SET
.
-
Test Analog Output
is shown.
2.
Press
SET
and choose between
Analog output 1, 2, 3
with
X
or
S
.
3.
Press
SET
, scroll one digit to the right using
X
and increase
value of digit by 1 using
S
. Any analog output value can be
predefined, e.g. for an analog output of 4 to 20 mA, the value
"6.0 mA". Confirm with
SET
or abort entry with
ESC
.
4.
Accept setting by pressing
SET
and test with multimeter
(minimum requirement: resolution 6.5 digits, at least 2-times
more accurate than the 6381) as follows:
Analog output 1 or 2: Via test contacts under service flap, see
diagram.
1
Channel 1
test contacts
2
Service
interface
3
Channel 2
test contacts
4
Ends of
channel 3
cables
5
Multimeter
Analog output 3: Connect measuring cables to terminals for
channel 3, guide measuring cables out of housing and perform
measurement outside of the transmitter, see diagram.
6.
Return to
Test Analog Output
with
ESC
and use
X
or
S
to
continue to Test Relay Output.
1.800.561.8187
www.
Testo-Direct
.ca