8 Tips and assistance 

18

 

 

 

8

 

Tips and assistance 

 

8.1.

 

Questions and answers 

Question 

Possible causes/solutions 

 flashes and 

E145

 

lights up 

The instrument temperature is outside 

the permissible range: 
>  Let the instrument warm up or cool 

down.  

 flashes and 

E161

 

lights up 
or 

 flashes and 

E419

 

lights up 
or 

 flashes and 

E999

 

lights up 

A serious error has occurred: 
>  Contact your dealer or the Testo 

Customer Service. 

 flashes and 

E290

 

lights up 

The current CO

2

 measuring value is too 

high for an adjustment: 
Acknowledge the error message with 

[Hold/Max/Min]

 and carry out the 

adjustment again in fresh air (CO

2

 

measuring value < 650 ppm). 

 flashes and 

E420

 

lights up 

Fan is blocked 
>  Please check if any item sticks in 

the ventilation slot on the back of 

the instrument. After 90 seconds, 

the fan is started again. 

>  If there is nothing directly visible, 

please contact your dealer or the 

Testo Customer Service. 

calibration

 and 

CO

2

 flash 

CO

2

 sensor must be adjusted 

>  Adjust CO

2

 sensor in fresh air. 

 flashes 

Battery capacity too low 
>  Connect the mains unit. 

Summary of Contents for 315-3

Page 1: ...testo 315 3 CO CO2 measuring instrument Instruction manual...

Page 2: ...2...

Page 3: ...er instrument data 7 3 2 4 Bluetooth module option 7 4 Product description 9 4 1 Overview 9 4 1 1 Control elements and connections 9 4 1 2 Display 12 4 2 Basic properties 13 5 First steps 13 6 Using t...

Page 4: ...writing standards Represen tation Explanation Note Basic or further information 1 2 Action more steps the sequence must be followed Action a step or an optional step Result of an action Menu Elements...

Page 5: ...mperatures For products with Bluetooth optional Changes or modifications that have been made without the explicit consent of the responsible approval authority may cause the retraction of the type app...

Page 6: ...ndustry The Bluetooth option may only be operated in countries in which it is type approved 3 2 Technical data 3 2 1 Measurement ranges and resolution Measurement parameter Measuring range Resolution...

Page 7: ...to 95 RH Weight incl holster Approx 382 g Housing material ABS Dimensions L x W x H 190 x 65 x 40 mm Protection class IP 40 acc to EN 60529 Operating time Min 10 h measuring time at 20 C 68 F mains o...

Page 8: ...inland FI France FR Greece GR Ireland IE Italy IT Latvia LV Lithuania LT Luxembourg LU Malta MT Netherlands NL Austria AT Poland PL Portugal PT Romania RO Sweden SE Slovakia SK Slovenia SI Spain ES Cz...

Page 9: ...4 1 Overview 4 1 1 Control elements and connections 1 Connection for temperature humidity module optional 2 IR interface record printer 3 CO sensor module 4 CO2 sensor module back 5 Display 6 Keyboar...

Page 10: ...figuration mode Confirms entry next setting Long Reading display current value Opens configuration mode Reading display maximum value minimum value hold value Resets maximum value minimum value hold v...

Page 11: ...n mode Changes option increases value Long Reading display Sets zero point deletes zero point Configuration mode Increases value quickly Short Reading display Prints data Configuration mode Changes op...

Page 12: ...abled Lights up information Flashes error message Differential measurement calibration Flashes adjust CO2 sensor in fresh air CO2 sensor should be adjusted in fresh air after 120 operating hours 2 Mea...

Page 13: ...45 C 32 to 113 F 5 First steps Switching on Connect the humidity temperature module 0636 9725 before switching the instrument on option 1 Switch the instrument on press briefly All segments of the di...

Page 14: ...cancel the function press briefly until the display changes Adjustable functions Ensure correct settings all settings are transferred immediately There is no Cancel function Settings and display of t...

Page 15: ...e carried out on busy roads or in closed rooms Ensure that before and during the calibration adjustment no exhaled air reaches the instrument Before completing the adjustment the instrument should be...

Page 16: ...Hold Max Min testo 315 3 is in measurement view The measuring values for Hold Max Min can be called up for CO and CO2 Temperature and humidity if temperature and humidity module is connected Dew poin...

Page 17: ...en carried out Switch on data transfer press and hold down lights up Data is transferred Please refer to the operating instructions for the relevant testo measuring instrument 7 Maintaining the produc...

Page 18: ...and E290 lights up The current CO2 measuring value is too high for an adjustment Acknowledge the error message with Hold Max Min and carry out the adjustment again in fresh air CO2 measuring value 65...

Page 19: ...ew point measurements be carried out Humidity temperature and dew point measurements are possible with the temperature humidity module accessory The display is illuminated also after the testo 315 3 i...

Page 20: ...ts Description Article no testo 315 3 without Bluetooth incl micro USB mains unit cable USB A USB micro B 0632 3153 testo 315 3 with Bluetooth incl micro USB mains unit cable USB A USB micro B 0632 31...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...0970 3153 en 04...

Reviews: