5 Parameterizing, adjusting and analyzing software (P2A software)
59
5.3.3.1.
Analyzing/testing the instrument
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Perform action:
Action
Explanation
Carrying out factory
reset
Reset the unit, limit value and hysteresis
parameters to factory settings (see below).
Testing analog
output
Test channel 1 (see Testing functionality of
analog outputs, page 34).
Test switch outputs Manually switch relays 1 to 4 to test for
proper function (see Testing functionality of
relay outputs, page 35).
Display min./max.
values
Overview of the minimum and maximum
values measured since the last reset of the
transmitter (see Displaying min./max. values,
page 62).
3.
Click on
[OK]
or
[Cancel]
to close the dialogue.
Pos: 92 /TD/Produkt verwenden/MUF 63xx/P2A/Werksreset durchführen @ 3\mod_1237379864072_79.docx @ 29887 @ 4 @ 1
5.3.3.2.
Carrying out factory reset
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark transmitter test.
- Current operating hours are shown.
3.
Confirm control query to perform the reset.
- The values are reset to the customer-specific factory settings.
4.
Click on
[OK]
or
[Cancel]
to close the dialogue.
Pos: 93 /TD/Produkt verwenden/MUF 63xx/P2A/Analogausgang testen 635x @ 3\mod_1237382683201_79.docx @ 29948 @ 4 @ 1
5.3.3.3.
Testing analog output
✓
The required instrument file is marked.
1.
Click on
[Test/analyze transmitter]
.
- The
Properties of <Instrument type> <Serial number>
dialogue is opened with the
Test/analyze transmitter
register.
2.
Mark channel and test values.
Summary of Contents for 0555 6383