4 Transmitter
41
Pos: 64 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Warnmeldungen 635x @ 3\mod_1236081573428_79.docx @ 27192 @ 3 @ 1
4.5.2.
Warning messages
Warning messages show an early warning or a current malfunction
which may negatively impact measuring.
Message Display
Cause
Remedying of fault
00E00
T ambient high The ambient
temperature exceeds
the permissible
temperature for the
transmitter
Take necessary
measures to lower
ambient temperature,
e.g. through venting or
cooling
00E01
T ambient low
The ambient
temperature is below
the permissible
temperature for the
transmitter
Take necessary
measures to raise
ambient temperature,
e.g. through heating
00809
Pressure too
high
The process pressure
exceeds the pressure
intended for the
transmitter
Remove the transmitter
from the process and
take any necessary
measures to lower the
pressure
00E02/00E
05
Supply voltage
low
The supply voltage is
below the minimum
voltage required for the
transmitter
Take measures to
ensure sufficient
voltage supply
0081C Alarm
1
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081D Alarm
2
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081E Alarm
3
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
0081F Alarm
4
Depending on the
parameterization of the
relay
Depending on the
parameterization of the
relay
Pos: 65 /TD/Produkt verwenden/MUF 63xx/Status-Warnmeldungen/Fehlermeldungen Messumformer 635x @ 3\mod_1236081613819_79.docx @ 27212 @ 3 @ 1
4.5.3.
Transmitter error messages
Error messages show a current malfunction.
Message Display
Cause
Remedying of fault