![TESTING 1.0334 Operating Manual Download Page 24](http://html1.mh-extra.com/html/testing/1-0334/1-0334_operating-manual_1092047024.webp)
Air Entrainment Meters
0.5 litre; model: TESTING 1.0334 0,75 litre; model: TESTING 1.0337
1 litre; model: TESTING 1.0335 5 litre; model: TESTING 2.0332
7 litre; model: TESTING 2.0337 8 litre; model: TESTING 2.0334
Page 24 of 42 pages
Note
In the event that the air entrainment meter does not show the required
5 % or 10 %, check the initial pressure once again. If necessary
send the complete air entrainment meter back to the manufacturer
for checking.
16. After taking the reading, slowly open the right ball shut-off valve (B) to allow the
pressure to be released from the test pot.
17. Open the quick-release latches and take off the upper part of the air entrainment
meter.
18. Open the overflow valve by pressing the TEST button and allow the rest of the air to
flow out of the pressure chamber. The pressure-gauge pointer will now go back slowly
and will reach its initial position. See Section 7. Calibration instructions, image 4.
19. Remove the calibration tubes. Empty the test pot and whip out the air entrainment
meter until it is dry.
20. Your air entrainment meter is now re-calibrated and is ready for operation.
5. Troubleshooting
Fault
Possible cause
Action to take
The air entrainment
meter repeatedly
shows incorrect
measuring results
The air entrainment meter has a
leak.
Get in touch with the manufacturer.
The pump has a leak.
The air entrainment meter must be
repaired. For this repair, get in touch with
the manufacturer.
Water or material to be tested is
in the pressure chamber.
The quick-release latches do not
close properly.
Adjust the quick-release latches. Then
recalibrate the air entrainment meter.
The pointer is not in correct
position.
Correct the position of the pointer. See
Section 4.1a.
The air entrainment
meter cannot be
pumped up.
The pump rod is bent.
The pump must be repaired.
The pump is clogged with dirt or
debris.
The pump must be repaired.
The pump sleeve is worn out.
The pump must be repaired.