background image

  
Air Entrainment Meters  

0.5   litre; model: TESTING 1.0334          0,75  litre; model: TESTING 1.0337 

 

1      litre; model: TESTING 1.0335           5  litre; model: TESTING 2.0332 

7      litre; model: TESTING 2.0337           8  litre; model: TESTING 2.0334 

 

  

  

Page 14 of 42 pages  

  
   

  

2.   The test procedure  

  

Caution  

 

Before filling the material to be tested into the test pot, make sure that 
all the inner surfaces of the test pot and the cover of this pot have been 
cleaned of oils, grease, fats, dust, and residue from earlier testing. Then 
slightly moisten the inner surfaces of the test pot with a moistened cloth 
or sponge. Only then may the new material to be tested be placed in the 
test pot.  

  
1.  Take  off  the  upper  part  by  unfastening  the  quick-release  latches.  Set  in  place  the 

attachment ring for filling the meter (this is an optional piece of equipment:  you may 
not have received this part). Then use the quick-release latches to connect the test 
pot and the attachment ring.  

2.  Fill the test pot with fresh mortar or fresh concrete in one or several layers, depending 

on their consistency and the stipulations of the relevant standard. Compact the mortar 
or concrete according to the relevant instructions of the standard by using a tamping 
rod,  other  rod,  or  internal  vibrator,  or  by  placing  the  air  entrainment  meter  on  a 
vibrating table. Perform the compacting in accordance with the instructions given in 
EN 

123507 

(

Luftgehaltsdruckausgleichsverfahren

 

air-content 

pressure-

compensation  method),  or  as  specified  by  other  standards  or  specifications.  The 
mortar  or  concrete must  be  completely  compacted.  If mechanical vibration  is  used, 
complete  compaction  is  achieved  when  large  air  bubbles  no  longer  appear  on  the 
surface,  when  the  surface  is  relatively  smooth,  and  when  it  appears  level.  If  an 
attachment ring is not used, the amount of material filled into the test pot must be such 
that it can be cleanly struck off with a straightedge.   

3.  After  compaction,  take  off  the  attachment  ring  (if  you  have  used  one)  and  use  a 

straightedge to strike off the concrete or mortar level with the upper edge of the test 
pot. Then use the smoothing trowel to smooth off the top surface.  

4.  Use a moist cloth or sponge to wipe clean the outer edges of the test pot, the flange, 

and the connection elements.  

5.  Moisten the inside of the upper part and set it down into place. For the 5-litre,7-litre 

and 8-litre air entrainment meters, turn the upper part until its red markings are in line 
with the markings on the test pot.  

6.  Lock downward the two quick-release latches, mounted opposite each other, to form 

an airtight seal between the upper part and the test pot.   

  
  

  

Important: correct positioning of the quick-release latches 

 

  

Summary of Contents for 1.0334

Page 1: ...struction material testing devices Motzener Str 26b DE 12277 Berlin Germany Tel 49 0 30 710 964 5 0 Fax 49 0 30 710 964 5 98 www testing de Operating Manual Air Entrainment Meter 0 5 litre 0 75 litre...

Page 2: ...ration of the system 10 1 6 2 Basic elements and characteristics of the air entrainment meters 11 2 The test procedure 14 3 Cleaning and maintenance 17 4 Checking and calibrating the air entrainment 1...

Page 3: ...attention to danger that can lead to damage to equipment or other objects Danger This warning indicates danger that can lead to severe injury or death Note This warning gives practical instructions o...

Page 4: ...hen this is authorized only after consultation with the manufacturer and after obtaining his prior and express approval The 0 5 litre air entrainment meter has been manufactured in accordance with the...

Page 5: ...air entrainment meters are damaged or have any malfunctions that can impair their operational safety they must be immediately placed out of operation and the malfunctions must be corrected These syste...

Page 6: ...f the meters Meter Weight when empty Weight when filled with mortar 0 5 litre air entrainment meter approx 3 8 kg approx 4 8 kg 0 75 litre air entrainment meter approx 3 9 kg approx 5 3 kg 1 litre air...

Page 7: ...t you return the system to us for repair 1 4 2 Transport This system will be delivered in the specific cardboard boxes In order to prevent transport damage the remaining hollow spaces in the interior...

Page 8: ...on Article number for the model 0 5 litre TESTING Article number for the model 0 75 litre TESTING Article number for the model 1 litre TESTING Article number for the model 5 litre TESTING Article numb...

Page 9: ...truction sites 1 0335 02 1 0335 02 1 0335 02 2 0333 25 2 0333 25 2 0333 25 Straightedge for striking off 400 mm 2 0237 04 Aluminium poking rod 2 0237 03 Rubber hammer 400 g 8 0204 Rubber hammer 800 g...

Page 10: ...se air entrainment meters have a pressure chamber in which a specified air pressure is produced by a hand pump When an overflow valve is opened pressure equalization is established with respect to the...

Page 11: ...ainment meters The air entrainment meters consist of 2 main components I Upper part a Cap pressure gauge mount Switch panel with integrated manual pump TEST valve correction valve and pressure gauge b...

Page 12: ...latches The water is filled in through the two ball shut off valves and air pressure release takes place through these same valves The compressed air is produced by a hand pump installed in the air en...

Page 13: ...mm 180 dia x 316 mm 180 dia x 330 mm 220 dia x 490 mm 220 x 505 mm 220 dia x 600 mm Net weight 3 8 kg 3 9 kg 4 0 kg 8 7 kg 9 1 kg 9 2 kg Gross weight approx 4 8 kg approx 5 3 kg approx 6 0 kg approx 2...

Page 14: ...by placing the air entrainment meter on a vibrating table Perform the compacting in accordance with the instructions given in EN 123507 Luftgehaltsdruckausgleichsverfahren air content pressure compens...

Page 15: ...and 1 litre TESTING air entrainment meter Adjust by making one complete turn of the hooks of the clamps For the 5 litre 7 litre and 8 litre TESTING air entrainment meters In making precision adjustmen...

Page 16: ...ave caused the black pointer to come to rest above the red marking next to the zero point In this case pump up the air entrainment meter again until the black pointer is below the red marking Note Aft...

Page 17: ...oncrete may get into the pressure chamber After pressure compensation the CORRECTION button must not be pressed under any circumstances as otherwise sample material will flow through the channels into...

Page 18: ...evice The initial pressure that applies to each device has already been set and marked by the red marking pointer during factory testing before each air entrainment meter is shipped to its users For p...

Page 19: ...r contains no more bubbles it is essential that the water continuously flows through the air entrainment meter 7 Now close both of the ball shut off valves A and B 8 Turn the ball headed knob to the l...

Page 20: ...t in touch with After Sales Service of the manufacturer His company will have to test and repair the air entrainment meter as necessary 4 3 Calibration From time to time it is necessary to check the a...

Page 21: ...capacity of the test pot by measuring the mass of water that it contains 2 First weigh the test pot together with a transparent plate that covers the top surface of the pot 3 Then fill the pot up to t...

Page 22: ...g or 100 g or 250 g or 350 g or 400 g of water 11 Carefully open the right ball shut off valve B to allow the pressure to be released from the test pot 12 Close both the ball shut off valves 13 Use t...

Page 23: ...00 750 g 1 litre TESTING model no 1 0335 Since 100 g of water was taken out of the test pot and the pot contains a water mass of 1000 g this amount of water removed is 10 by volume This value must be...

Page 24: ...nitial position See Section 7 Calibration instructions image 4 19 Remove the calibration tubes Empty the test pot and whip out the air entrainment meter until it is dry 20 Your air entrainment meter i...

Page 25: ...he top of the test pot Follow the instructions for the use of the air entrainment meter 6 After sales service and spare parts A great deal of care has been taken to ensure that this Operating Manual i...

Page 26: ...Feuerherdt GmbH Motzener Str 26b DE 12277 Berlin Germany Tel 49 30 710 96 450 Telefax 49 30 710 96 45 98 www testing de info testing de 6 4 Disposal and recycling These air entrainment meters be disp...

Page 27: ...1 0335 A2 02 2 Quick release latch complete 2 x screws each 10 1 0335 A4 1 Clamping stud complete O rings nuts tube fitting 11 1 0335 A6 1 Pressure gauge complete Tube fitting O ring 12 1 0335 A7 K 1...

Page 28: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 28 o...

Page 29: ...1 0335 A2 02 2 Quick release latch complete 2 x screws each 10 1 0335 A4 1 Clamping stud complete O rings nuts tube fitting 11 1 0335 A6 1 Pressure gauge complete Tube fitting O ring 12 1 0335 A7 K 1...

Page 30: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 30 o...

Page 31: ...0335 A2 02 2 Quick release latch complete 2 x screws each 10 1 0335 A4 1 Clamping stud complete O rings nuts tube fitting 11 1 0335 A6 1 Pressure gauge complete Tube fitting O ring 12 1 0335 A7 K 1 H...

Page 32: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 32 o...

Page 33: ...uick release latch complete 2 ea screws and spring 10 2 0334 A4 1 Clamping stud complete O rings nuts hose fittings 11 2 0332 A6 1 Pressure gauge Tube fittings O ring 12 2 0334 MS 1 Spare parts kit fo...

Page 34: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 34 o...

Page 35: ...uick release latch complete 2 ea screws and spring 10 2 0334 A4 1 Clamping stud complete O rings nuts hose fittings 11 2 0337 A6 1 Pressure gauge Tube fittings O ring 12 2 0334 MS 1 Spare parts kit fo...

Page 36: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 36 o...

Page 37: ...k release latch complete 2 ea screws and spring 10 2 0334 A4 1 Clamping stud complete O rings nuts hose fittings 11 2 0334 A6 1 Pressure gauge Tube fittings O ring 12 2 0334 MS 1 Spare parts kit for t...

Page 38: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 38 o...

Page 39: ...5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 39 of 42 pages 7...

Page 40: ...ment Meters 0 5 litre model TESTING 1 0334 0 75 litre model TESTING 1 0337 1 litre model TESTING 1 0335 5 litre model TESTING 2 0332 7 litre model TESTING 2 0337 8 litre model TESTING 2 0334 Page 40 o...

Page 41: ...ed past the red marking Please do not pump any further Too much pressure may cause damages inside of the device or at the pressure gauge After the operator has pressed the Test button he or she must n...

Page 42: ...ed guidelines as well as the additional applied guidelines following including any of the amendments thereto which are in force at the time of the declaration The following additional EU Directives ha...

Reviews: