Chapter 6 – Specifications
TestEquity 123HS Temperature/Humidity Chamber
Page 6-1
Chapter 6 – Specifications
Temperature Range
–35°C to +175°C
Control Tolerance
±0.5°C (Short-term variations measured at the control sensor after stabilization)
Uniformity
±1.0°C (Variations throughout the chamber after stabilization)
Standard Humidity Range
10% to 95%, in the temperature range of 10°C to 85°C limited by a 6°C dewpoint;
See chart on page 4-6.
Control Tolerance
±3%RH (Short-term variations measured at the control sensor after stabilization)
Live Load Capacity @
+23°C
0°C
–10°C
–20°C
–30°C
500 W
400 W
300 W
200 W
100 W
(Live load capacity is for non-humidity mode)
Cool Down Transition Time*
End Temp
Start Temp to
+23°C
0°C
–10°C
–20°C
–30°C
–35°C
+23°C
----
4 min
8 min
15 min
25 min ultimate
+50°C
5 min
10 min
16 min
24 min
36 min ultimate
+85°C
10 min
19 min
25 min
33 min
45 min ultimate
+150°C
22 min
33 min
40 min
48 min
63 min ultimate
Heat Up Transition Time*
End Temp
Start Temp to
+23°C
+50°C
+85°C +125°C +150°C +175°C
+23°C
-----
2 min
12 min
24 min
33 min ultimate
0°C
3 min
9 min
18 min
32 min
41 min ultimate
–30°C
8 min
14 min
23 min
27 min
45 min ultimate
*Note:
Transition times are measured after a 2 hour soak at the respective start temperature with an empty chamber,
as indicated on the temperature controller, 23°C ambient. Measured with setpoint beyond the start and end
temperatures. Does not include the effect of proportional band when approaching setpoint.
Input Power
120 V nominal (110 to 126 VAC), 60 Hz, 1 PH
Max Current Draw 18 A, Recommended Minimum Service 20 A (dedicated line)
Workspace Dimensions
18" W x 16.5" H x 13.5" D (2.3 cubic feet)
Outside Dimensions
26" W x 63" H x 36.5" D (nominal). Vent tube adds 3" to height.
Min. Installed Clearance
6" from the left and right side, 12" from the rear
Access Ports
4" Port on left and right side (two total), Supplied with foam plugs
Sound Level
58 dBA (A-weighted, measured 36" from the front)
Weight
750 pounds
NOTE:
Performance is typical and based on operation at 23°C (73°F) ambient and nominal input voltage. Designed
for use in a normal conditioned laboratory. Operation at higher ambient temperatures may result in decreased
cooling performance. Additional ports and shelves will also affect performance. Operation above 30°C (85°F) or
below 16°C (60°F) ambient is not recommended.
Due to continuous product development, specifications are subject to change without notice.
Summary of Contents for 123HS
Page 2: ......
Page 20: ......
Page 24: ......
Page 28: ......
Page 30: ......
Page 48: ......
Page 50: ......
Page 51: ...Chapter 9 Drawings TestEquity 123HS Temperature Humidity Chamber Page 9 1 Chapter 9 Drawings...
Page 52: ......
Page 53: ...Electrical Subpanel Component Location Purge Subpanel optional Main Subpanel...