![Test Instrument Solutions SafetyPAT Operating Manual Download Page 17](http://html1.mh-extra.com/html/test-instrument-solutions/safetypat/safetypat_operating-manual_1091886017.webp)
16
12. Environmental Condition
Operating temperature range 0
℃
to 40
℃
,
without moisture condensation
。
Storage temperature range 0 to 40C
Batteries must be removed prior to storage
。
13. Main Technical Parameters
Earth Continuity Range 0.00
Ω
to 19.99
Ω
Pass Limit
0.01
Ω
-0.1
Ω
solid Pass Only
0.1
Ω
- 0.5
Ω
flashing Pass Only
0.51
Ω
- 19.99
Ω
shows Fail Only
Test Current
Resistance
≤
2 ohms
:
≥
200mA
Test Voltage
5V DC
Test Time 5 Seconds