background image

 

3/8 

 

TSR4 is designed for continuous measuring of radon concentrations in buildings. 
Portable  probe  basis  is  a  measuring  chamber  with  a  semiconductor  photodetector.  Radon  enters  the 

chamber by diffusion through the input filter on the bottom of probe. The probe measures in autonomous 
and  time  continuous  way.  It  processes  results  every  4-minute  intervals  and  from  this  counts  short-term 
moving  average  of  radon  concentration  (  1  hour  moving  average  -  average  of  15  4-minute  process 
intervals). It also counts long-term moving average of radon concentration (24 hours moving average). The 
probe saves time records of these radon concentration values including values of humidity and temperature 
within its internal memory (typically at an interval of 1 hour). Next saved value is time record of measuring 
energy spectrum (typically at an interval of 12 hours). The probe is random for location in measured place, 
but generally it is put on the bottom of the probe. Bottom of the probe cannot be covered. The probe can 
be  switched  on/off  by  switch.  LEDs  „STAT“  and  „CHRG“  indicate  current  status  of  probe  see    ´Operation 
manual´ below. Waterproof cover for electronics (IP67). 

The  resulting  values  can  be  downloaded  continuously  during  measurement  or  at  once  at  the  end  of 

measurement.  Data  from  the  radon  probe  can  be  downloaded  to  a  PC  directly  via  USB  interface  or 
wirelessly via antenna and Central Unit. Central Unit is not included with package of TSR4 and it is sold and 
delivered extra ; see

 

http://www.tesla.cz/

 

.

 

 

TSR4A Radon Probe can be operated by these ways: 

 

A)

 

Standalone  probe

  -  Thanks  to  its  independent  battery  power,  portable  radon  measuring  probe 

supports flexible placing options within monitored structures. Accumulator will last for more than 
1 year after full charging. After switching on TSR4 immediately starts measuring and saving results 
into internal memory. The resulting values are downloaded after end of the measurement by B) or 
C) way. 

 

 

B)

 

Probe  connected  via  USB

  –  Using  usbRADONview  app  and  USB  cable  is  possible  to  download 

results  to  PC  continuously  during  measurement  or  at  once  at  the  end  of  measurement. 
usbRADONview  application,  drivers  and  user  manual  is  free  downloaded  on  website: 

http://www.tesla.cz/

. 

 

 

C1)   

Wireless network for radon measurement

 - Central Unit supports simultaneous data downloads 

from  up  to  16  probes.  All  values  are  saved  again  into  central  unit  memory.  Via  connected 
computer to Central Unit and TERAview application on PC it is possible to download and process 
all  data  from  system  and  configure  whole  system.    Setting  and  configuration  of  TSR4A  Radon 
Probe and whole system is also managed by TERAview application on PC. TERAview application 
version  3.11.6  and  higher,  drivers  and  user  manual  with  detail  configuration  description  is  free 
downloaded on website

http://www.tesla.cz/

. 

   In  case  of  time  continuous  measurement  of  radon  concentration  or  in  case  of  setting  in 

regulation system the probe must be placed in radio range of Central Unit. Distance (radio range) 
between TSR4 and central unit is up to 600m in open space. In buildings it depends on number of 
walls, building material, etc. Strength of radio signals (RSSI) is monitored by Central Unit. If radio 

Summary of Contents for TSR4A

Page 1: ...Specifications Operation Manual v 1 2022 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 5 4 Product Specification 6 5 Operating Instructions 6 6 Basic Maintenan...

Page 2: ...used product If you experience any problems with your product please contact us at TESLA Rubeska 215 1 190 00 Prague 9 Vysocany www tesla cz 1 Introduction This document describes technical specifica...

Page 3: ...sla cz TSR4A Radon Probe can be operated by these ways A Standalone probe Thanks to its independent battery power portable radon measuring probe supports flexible placing options within monitored stru...

Page 4: ...centration values to central unit wirelessly Central unit analyzes this information and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuato...

Page 5: ...5 8 3 Scope of Delivery TSR4A WLAN Wireless Radon Probe Power adapter 230VAC 5VDC USB cable A B Antenna Probe holder Operation Manual...

Page 6: ...on operating conditions Current radon concentration short term 0 5 hour running average from RaA long term 24 hours running average from RaA RaC Dimension 80 x 175 mm Waterproof IP67 only for electron...

Page 7: ...way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application UsbRADONview appl...

Page 8: ...recalibration is free from the manufacturer 7 EC Declaration of Conformity EC Declaration of Conformity will be delivered by Tesla producer on request If interested please use contacts on the web www...

Reviews: