LED diode „STAT“:
It signalizes status radon probe according to following chart:
Color
Description
Green blink 3x
Radon probe has just been turned on.
Green blink after 8s
Radon probe measures and works correctly
Yellow blink 4x
Radon probe has just been turned off.
Green
/
Yellow
blink after 5s
Radon probe measures but troubles are occur. – especially low capacity of
accumulator. Warnings and errors are shown in PC application.
No light, No blinking
Radon probe doesn´t measure or accumulator is empty or device is damaged.
Charging process of accumulator is described in chapter „Basic Maintenance/
Accumulator charging“
Antenna installation:
Supplied antenna is screwed on antenna connector. When installing antenna, hold antenna by knurled
end.
Power supply:
According to operation method the radon probe can be supplied:
1)
By internal accumulator for portable use – Radon probe includes internal accumulator which is
able to ensure autonomous operation of probe for more than 12 months without charging.
Depends on frequency of wireless data downloading from probe and depends on climatic
condition of probe use. Accumulator is charged over USB port and over provided USB cable. The
USB cable is possible to connect to PC or to delivered power adapter. Status of accumulator and
charging process are described in paragraph ´Basic Maintenance/Accumulator charging´
2)
By mains power supply 230V/50Hz for stationary use – Radon probe is permanently supplied by
delivered power adapter. Power adapter is connected to probe via provided USB cable. In case of
blackout internal accumulator ensures UPS function.
Configuration:
Setting and configuration is different according to operation way of radon probe.
If the probe is used for autonomous measurement and data downloading via USB interface then the
setting and configuration are realized by usbRADONview application. UsbRADONview application, drivers
and user manual with detail configuration description are free downloaded on website:
If the probe is used for wireless measurement and wireless data downloading then setting and
configuration of probe and whole measuring system is realized by Central Unit connected to PC and
TERAview application.
Central Unit is not included with package of probe and it is sold and delivered extra;
see
. TERAview application, drivers and user manual with detail configuration
description are free downloaded on website:
. For successful probe configuration in
measuring system is essential to know probe radio channel number (communication wireless channel) and
P2P address (identification in wireless net). Both parameters are printed out on probe serial number plate.
Probe radio channel number is possible to change by TERAview application and it must be identical to
central unit radio channel number. P2P address is permanent and it can occur in one big wireless net only
once. Only Central Unit P2P address can be identical to P2P address of other elements in network.
6
Basic Maintenance
Accumulator charging:
During portable use of radon probe is essential to monitor state of internal accumulator and recharge it if
necessary. If accumulator is discharged the probe automatically turns off. The probe is switched on again
powering USB port.
Current state of accumulator can be determined by these ways:
1) By LED diode ´STAT´ - If LED starts blinking in green-yellow color it indicates that system is working
incorrectly and one of main case is low voltage of accumulator. (see paragraph “Operation Manual /
LED diode “STAT””)