EUT power ON/OFF function 

Front panel switch with ON indicator

Fuse 

16 A, slow blow

Connectors 

Harting type HAN3A compatible with 

 

 

NSG 3000 safe linking concept

Possible extensions 

Interlock connection to door switch

Own supply 

Selectable 100-110 V; 220-240 V 15 VA

Dimensions:  

L x W x H 460 x 200 x 160 mm

 

 

(18.1 x 7.9 x 6.3”)

Weight 

15 kg (33 lbs) approx.

Cable length to NSG 3000 

2 m (79”)

Input cable 

NSG 3000 standard cable to be used

Control cable 

2 m - 25 way sub D – twisted pair - 

 

 

shielded (included in delivery)

 

Part designation 

Function

Mains supply in socket 

For instrument supply, includes plug, 

 

 

voltage selector (110 V – 230 VAC), on/off 

 

 

switch and fuse

EUT power in switch 

Switches the EUT power supply on/off

16 AT fuse 

Protects the EUT power supply line

EUT power in plug 

For feeding of the EUT power – use NSG 

 

 

supplied cable and connect it to power 

 

 

source (mains)

EUT power out cable 

Connect this cable to the EUT supply 

 

 

input port of the NSG

 

2.2.1 Parts description

Summary of Contents for INA 6501

Page 1: ...1 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 USER MANUAL 601 262C...

Page 2: ...Manual automated step transformer INA 6501 INA 6502 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 601 262C USER MANUAL...

Page 3: ...1 2 Technical specifications INA 6501 6 1 2 1 Parts description 7 1 3 Installation 7 2 Automated step transformer INA 6502 9 2 1 Circuit diagram INA 6502 10 2 2 Technical specifications INA 6502 10 2...

Page 4: ......

Page 5: ...in case of use in standing posi tion as the stability is limited The cabling connecting INA 6501 to mains and NSG presents a risk of being unvoluntarly caught by the users which could cause the INA 6...

Page 6: ...0 A less than 5 Output current capability at 16 Arms at 100 output 230 V input voltage 20 Arms at 80 output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection Front pa...

Page 7: ...is where the EUT power needs to be applied use NSG 3000 supplied cable and conect it to power source mains EUT power out cable Connect this cable to the EUT supply input port of NSG 3000 Red selector...

Page 8: ...00 generator Select the required variable voltage using the rotary switch on the INA 6501 Because of the capacitors in the internal coupler of the CDN earth leakage currents of up to 4 A can occur in...

Page 9: ...A 6502 is fitted with carrying handles as part of its overall excellent ergonomic design which allows easy handling Further the unit may be used in any of three operating positions laying or standing...

Page 10: ...output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection User interface or WIN 3000 software 2 1 Circuit diagram INA 6502 L 1 1 3 2 2 EUT Power IN N C PE N EUT Power...

Page 11: ...m 79 Input cable NSG 3000 standard cable to be used Control cable 2 m 25 way sub D twisted pair shielded included in delivery Part designation Function Mains supply in socket For instrument supply inc...

Page 12: ...out to be terminated by interlock plug or linked to another accessory to X1 plug Plug X1 NSG 3000 system interface in to be connected to master controller or to another accessory to X2 plug Power LED...

Page 13: ...mains using EUT power in cable delivered with NSG or CDN Because of the capacitors in the internal coupler of CDN earth leakage currents of up to 4 A can occur in the EUT power supply network The tes...

Page 14: ...nterprise Shanghai Co Ltd Beijing Branch T 86 10 8526 2111 F 86 10 8526 2141 chinasales teseq com France AMETEK SAS T 33 1 30 68 89 00 F 33 1 30 68 89 99 info france ametek com Germany Teseq GmbH T 49...

Reviews: