9

The automated step transformer type INA 6502 is a standard accessory for 
the Teseq NSG 3000 instrumentation series. It provides a convenient means 
of reducing the incoming supply voltage by pre-set amounts. It is required for 
power quality testing (PQT) and is fully compliant with the latest revision of IEC 
61000-4-11.

Its control is fully automatic, driven from the NSG 3000 master controller, using 
the touch screen interface or the WIN 3000 remote control software. Once 
detected, the functions offered by the INA 6502 are available in the interface or 
software. So the settings 0 - 40% - 70 - 80% will appear, as well as the possibility 
to switch the EUT power ON/OFF.

The INA 6502 is fi tted with carrying handles as part of its overall excellent 
ergonomic design, which allows easy handling. Further, the unit may be used in 
any of three operating positions; laying or standing on a workbench. For more 
permanent applications, it can be even wall-mounted.

The unit has been designed for use in rugged industrial environments. High 
quality connectors ensure user safety, additional system protection is provided 
by a 16 A fuse located in the top panel.

Thanks to the provision of an 80% voltage position and to the large over current 
capabilities the step transformer is fully compliant with the latest requirements 
called for in IEC 61000-4-11 standard.

2  AUTOMATED STEP TRANSFORMER 

INA 6502

 

For proper operation of the plug and play detection 
mechanisms it is strongly recommended to power on 

fi rst the INA 6502 accessory and then the NSG main 

frame. 

Summary of Contents for INA 6501

Page 1: ...1 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 USER MANUAL 601 262C...

Page 2: ...Manual automated step transformer INA 6501 INA 6502 MANUAL AUTOMATED STEP TRANSFORMER INA 6501 INA 6502 601 262C USER MANUAL...

Page 3: ...1 2 Technical specifications INA 6501 6 1 2 1 Parts description 7 1 3 Installation 7 2 Automated step transformer INA 6502 9 2 1 Circuit diagram INA 6502 10 2 2 Technical specifications INA 6502 10 2...

Page 4: ......

Page 5: ...in case of use in standing posi tion as the stability is limited The cabling connecting INA 6501 to mains and NSG presents a risk of being unvoluntarly caught by the users which could cause the INA 6...

Page 6: ...0 A less than 5 Output current capability at 16 Arms at 100 output 230 V input voltage 20 Arms at 80 output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection Front pa...

Page 7: ...is where the EUT power needs to be applied use NSG 3000 supplied cable and conect it to power source mains EUT power out cable Connect this cable to the EUT supply input port of NSG 3000 Red selector...

Page 8: ...00 generator Select the required variable voltage using the rotary switch on the INA 6501 Because of the capacitors in the internal coupler of the CDN earth leakage currents of up to 4 A can occur in...

Page 9: ...A 6502 is fitted with carrying handles as part of its overall excellent ergonomic design which allows easy handling Further the unit may be used in any of three operating positions laying or standing...

Page 10: ...output for 5 s 23 Arms at 70 output for 3 s 40 Arms at 40 output for 3 s Voltage selection User interface or WIN 3000 software 2 1 Circuit diagram INA 6502 L 1 1 3 2 2 EUT Power IN N C PE N EUT Power...

Page 11: ...m 79 Input cable NSG 3000 standard cable to be used Control cable 2 m 25 way sub D twisted pair shielded included in delivery Part designation Function Mains supply in socket For instrument supply inc...

Page 12: ...out to be terminated by interlock plug or linked to another accessory to X1 plug Plug X1 NSG 3000 system interface in to be connected to master controller or to another accessory to X2 plug Power LED...

Page 13: ...mains using EUT power in cable delivered with NSG or CDN Because of the capacitors in the internal coupler of CDN earth leakage currents of up to 4 A can occur in the EUT power supply network The tes...

Page 14: ...nterprise Shanghai Co Ltd Beijing Branch T 86 10 8526 2111 F 86 10 8526 2141 chinasales teseq com France AMETEK SAS T 33 1 30 68 89 00 F 33 1 30 68 89 99 info france ametek com Germany Teseq GmbH T 49...