46
Performance Test
6.2.3 Performance Test Set Up
Figure 6-2
Schematic Design for TC-5922BP Performance Test
6.2.4 Shielding Standards
Frequency
Shielding effectiveness (dB)
100 to 2000 MHz
> 70 dB
2000 to 3000 MHz
> 70 dB
3000 to 6000 MHz
> 60 dB
The shielding effectiveness above is measured with blank panels; other I/O interface panels
may result in different shielding effectiveness of the shield box. Please refer to
.
6.2.5 Test Procedure
Before measuring, check the shielding gasket around the lid. If there is a damaged part,
replace the gasket and wipe the contact area, which is between the lid and the body except
the gasket, with alcohol.