RS-232C Command List
33
5.5.1 Error Data List
Table 5-3
List of Error Data
Error Code
Description
ERR10
Syntax error. The command does not exist or is invalid.
ERR20
Handler-inward sensor error. It only occurs when the handler is in sensor mode.
(The horizontal operation of fixture is incomplete.)
ERR30
Probe-downward sensor error. It only occurs when the probe is in sensor mode.
(The vertical operation of fixture is incomplete.)
ERR40
Status error, It occurs when the command is not executable. For instance,
OUTWARD (handler) or UPWARD (probe) while the lid is closed. The error is
returned if a command is sent when the motion can not be done in conjunction
with the current status. It also occurs when only one button is pressed to close the
lid.
ERR50
Lid-open sensor error. It occurs only to certain products which should always detect
whether the lid is open. The response comes when the lid-open sensor does not
detect the main cylinder in open position.
ERR60
Safety error. The lid-close sensor does not detect the cylinder in close position.
ERR90
It occurs when a fixture operation command is sent while the corresponding
fixture is off. It also occurs when either one or both buttons are let go while closing
the lid.
Summary of Contents for TC-5922BP
Page 1: ...TC 5922BP Pneumatic Shield Box User Manual R20200211...
Page 9: ...PART 1 1 Getting Started...
Page 13: ...PART 2 5 Hardware Overview...
Page 16: ...8 Control Box 2 3 Control Box Figure 2 4 TC 5922BP Control Box Figure 2 5 5922BP CPU Board...
Page 18: ...10 Dimensions 2 4 Dimensions...
Page 19: ...Dimensions 11...
Page 24: ...PART 3 16 Putting into Operation...
Page 51: ...PART 4 43 Maintenance...