4. Technical Specifications
General
Case ........................................: ABS shielded for EMI / EMC compatibility.
Connections .............................: · "RTX" connectors IN/OUT and
“TX” OUT triple
banana-jack;
· polarized connector for external supply;
· RJ-45/4 connector for headset;
· USB port for PC connector and RS232 (to service);
· USB for pen-drive connector.
Display .....................................: LCD color 320 x 240 pixel (
1/4 VGA
) backlit.
Power supply (internal) .............: · rechargeable battery pack (green) NiMh, with life
of about 8 hours (typical), 5 hours (minimum).
·
external: from 16,5 to 26,5 Vdc / max 2,5 Ah.
Dimensions and Weight ............: 150 x 210 x 50 mm / 1,5 Kg. (batteries included).
Temperature range ...................: Operating: 0
+45° C. / Storage -20
+70° C.
Humidity range .........................: 5
90% non-condensing.
Overvoltage
Protection………… : In/Out RXT connectors up to 150 Vdc / 140 Vpp.
Reference Frequency accuracy : ·
1 ppm within the operating temperature
2 ppm/year.
Reference Level
accuracy…….. : ·
0.025% within the operating temp.
0.025% / year.
CE mark - EMC
…………………:
Directive 2004/108/CE, 89/336/EEC, Decree 2007/194
CISPR 11, ISO 14253 and CEI EN: 61326/A1/A2, 55011,
61000-4-2, 61000-4-3, 61000-4-4, 61000-4-6, 61000-4-11.
Special Features and Setup ......:
● Results storage on internal flash memory;
● Software update and Results Export on Pen Drive;
● PC Remote interface by USB port.
Level Generator
Sine output frequency range .....: · base band: from 20 Hz to 22 kHz;
· medium band: from 20 kHz to 6 MHz;
· high band (optional): from 20 kHz to 35 MHz.
Reference Frequency accuracy.: ·
1 ppm within the operating temp.
2 ppm/year.
Reference Level accuracy.........: ·
0.025% within the operating temp.
0.025% / year.
Frequency Resolution ...............: 1 Hz up to 9.999999 MHz; 10 Hz over 10.0 MHz.
Frequency setup mode .............: manual on single frequency and step mode on
programmable band / steps.
Balanced output impedances ....: · base band: 150, 200 and 600
Ω;
· medium band: 100, 110, 120, 135, 150
Ω;
· high band: 100
Ω.
Unbalanced output impedances: 50, 55, 60, 68 and 75
Ω by Banana/BNC optional adapt.
Output level
– Base Band……..: ·
-70
¸
+14 dBm @ 600
Ω balanced / 0.1 dB steps;
·
-64
¸
+17 dBm @ 75
Ω unbalanced / 0.1 dB steps.
Output level
– Medium Band.....: ·
-64
¸
+20 dBm @ 100
¸
150
Ω balanced / 0.1 dB steps;
·
-64
¸
+17 dBm @ 50
¸
75
Ω unbalanced / 0.1 dB steps;
Output level
– High Band (opt.) .: ·
0 dBm @ 100
Ω balanced;
Output level
– High Band (opt,) .: ·
0 dBm @ 50
Ω unbalanced.