ACT6000 User Guide
– January 2021 Page 47 of 52
13. File Manager
From the Main menu select the
“
6. File Manager
”.
All the measurements saved in the memory,
they can be seen and/or exported to the USB
Pen-drive or PC.
To view a measurement, select the file by the
arrows, then push
F5
“
View
”.
To cancel or to export one or more files
select the file by
and push the
F1
“
Select
”
key, repeat this procedure to select others
possible files.
Push the
F4
“
Export
” key to export the
selected file on the USB Pen-drive or PC (in
.CSV
*
format) or to cancel the file.
Pushing and selecting
“All files” by
F2
“
Delete
” key it is possible to erase the memory.
By the
F3
“
Sort
” key it is possible to arrange the data file for “
Date
” or “
Name
”.
Note* Opening an exported CSV file by a PC with Windows Excel, it is possible to view the test
parameters and all the row data results.
Selecting the columns related to the transmissive row data it is possible to convert these in graphics
useful to edit some possible Test Reports.
ACT6000 CSV-PDF Converter
By this PC Utility, free downloadable from
https://tescomusa.com/pages/act6000-
it is possible to convert automatically the in PDF the CSV files related to the
automatic
SELT
and
DELT
Sequences
performed thanks to the
ASW-1/II
(Advanced
Software), to produce the useful
Test Report
.
Note: Some transmissive threshold limits of the above PC Utility are similar to the limits set on the
instrument (standard version), vice-versa in some cases the limits for
DELT
and
SELT
Sequences
are
different to satisfy the pass/fail indications according to the indication displayed by the instrument.
The limits setup is listed on the
act6000.ini
present into the
“
TesCom_CSV_PDF_Converter
” PC
folder.
SELT Electric Tests
Vac A-B, Vac A-C, Vac B-C = <2.0 (Vac), Vdc A-B, Vdc A-C, Vdc B-C = <2.0 (Vdc)
Insulation (isol) A-B, A-C, B-C = >20 (MOhm)
SELT / DELT Transmissive tests
Noise (rumore)_Voice
<-47.0
(dBm)
SELT Ret-loss (local)
<-14
(dB) same for any mask
Noise (rumore)_Modem 56k
<-47.0
(dBm)
DELT Ret-loss local
<-14
(dB) same for any mask
Noise (rumore)_ISDN
<-45.0
(dBm)
DELT Ret-Loss remote
<-14
(dB) same for any mask
Noise (rumore)_HDSL1p
<-45.0
(dBm)
Bilan_long (local)
<-45
(dB) same for any mask
Noise (rumore)_HDSL2p
<-45.0
(dBm)
DELT Bilan-long (local)
<-45
(dB) same for any mask
Noise (rumore)_SHDSL
<-45.0
(dBm)
DELT Bilan-long (remote)
<-45
(dB) same for any mask
Noise (rumore)_E1 / T1
<-45.0
(dBm)
Ins-Loss (att)
Voice/Modem
DELT
>-11
(dB),
SELT
>-12
(dB)
Noise (rumore)_ADSL
<-45.0
(dBm)
Ins-Loss (att) ISDN
DELT
>-37
(dB),
SELT
>-38
(dB)