28
h
scanner sensors - if the scanner sensors are block by dust,
clean the sensor holes (highlight areas) using a cotton swab
moistened with the isopropyl alcohol.
Calibrating Scanner
When scanned image’s colors appear not to match the colors of the
original, or other colors appear in the white areas of the images,
calibration can help to correct these problems by compensating
variations in the scanner components that happens over time.
maintenance
1. turn on the power of the scanner.
2. remove the plastic foil on both sides of the calibration sheet.
3. insert the calibration sheet into the input tray with the arrow
mark facing up and forward.
4. Go to the start menu, select All Programs >
idocscan P series > calibration to open the
calibration dialog box.
5. click
Calibrate
to start calibrating. the
calibration sheet will be gradually fed
through the scanner.
h
A progress bar will appear indicating the calibration process.
h
mechanical sounds produced during the calibrating process are
normal.
6. click
Close
after “calibration done.” dialog box appears.