2. SPECIFICATIONS
2-1 General Information
Display :
Dual display LCD.
Measurement Range :
Type K :-190
℃
to 1333
℃
(-310
℉
to 2431
℉
)
Type J :-190
℃
to
760
℃
(-310
℉
to 1400
℉
)
Resolution :
0.1
℃
, 1
℃
, 0.1
℉
, 1
℉
.
Input Protection :
60Vdc, or 24 Vrms ac.
Sampling Rate :
1 time per second.
Memory capacity :
8000 records data, utmost 255 sets of results.
Over range indication :
“OL” appears on the display.
Break or no input indication : “- - - - -” appears on the display.
Low battery indication :
The is display when the battery voltage
drops below the operating voltage.
Power supply :
Single 9V battery 006P 9V or IEC 6F22, or
NEDA 1604.
Battery Life :
Approx. 35 hours (Alkaline battery).
Operating temperature
0
℃
to 40
℃
(32
℉
to 104
℉
) ,
and humidity : 10 – 80%RH
Storage temperature
-10
℃
to 60
℃
(14
℉
to 140
℉
) ,
and humidity : 10 – 70%RH
Dimensions :
145mm(L) x 68mm(W) x 35mm(H)
5.7(L) x 2.7(W) x 1.4(H) inches
Weight :
Approx. 235g with battery.
Accessories :
Instruction manual, carry case, battery,
CD Software, RS-232 cable, 9 pin to 25 pin
gender changer.
2
2-2 Electrical Specifications
Basic Accuracy: ( @23
±
5
℃
Calibration ) Accuracy are
±
( ...
%
of
r degree ) at 18
℃
to 28
℃
with relative humidity up to 80
%
.
For single thermocouple measurements
Function Resolution
Range
Accuracy
Type K
Type J
℃
0.1
℃
-190
℃~
-0
℃
-190
℃~
-0
℃
±
( 0.5%rdg + 0.7
℃
)
0
℃~
1000
℃
0
℃~
760
℃
±
( 0.1%rdg + 0.5
℃
)
1.0
℃
1000
℃~
1333
℃
±
( 0.2%rdg + 1
℃
)
℉
0.1
℉
-310
℉~
-0
℉
-310
℉~
-0
℉
±
( 0.5%rdg + 1.4
℉
)
0
℉~
999.0
℉
0
℉~
999.0
℉
±
( 0.1%rdg + 1.0
℉
)
1.0
℉
1000
℉~
2431
℉
1000
℉~
1400
℉
±
( 0.2%rdg + 2
℉
)
Temperature Coefficient:
0.1 times the applicable accuracy specification per
℃
from o
℃
to 18
℃
and 28
℃
to 40
℃
( 32
℉
to 64
℉
and 82
℉
to 104
℉
).
For T1-T2 Measurements accuracy is Basic accuracy add 0.2%rdg.
NOTE
The basic accuracy specification does not include the error of the probe.
Please refer to the probe accuracy specification for additional details.
3
+