For Medical Staff
Table c) – Immunity test level to proximity fields from RF communications equipment – for IEC 60601-1-2:2014 (EN 60601-1-2:2015)
Immunity test level to proximity fields from RF communications equipment
The TERUFUSION Syringe Pump Type SS3 is intended for use in the electromagnetic environment specified below. The
customer or the user of the TERUFUSION Syringe Pump Type SS3 should assure that it is used in such an environment.
Emissions test
Test frequency (MHz)
Immunity test level (V/m)
Proximity fields from RF communications
equipment
IEC 61000-4-3
385
27
450
28
710, 745, 780
9
810, 870, 930
28
1720, 1845, 1970
28
2100
10
2450
28
2600, 3500
10
5240, 5500, 5785
9