HARDWARE SET-UP
The SmartPReP 2 System is packaged in one box. Verify that the box contains one (1) SmartPReP 2
centrifuge, line cord, and Operator’s and Service Manual.
NOTE:
Save box, packaging and rotor support foam for future use.
1.
Remove the SmartPReP 2 centrifuge from the box and place on a sturdy counter or stable cart.
Remove all shipping restraints and inspect for damage.
2.
Install line cord and plug into the appropriate mains voltage.
3.
Visually inspect the display panel and verify the Green “Power” light and the yellow “Interrupted”
light are illuminated. Press the LID keypad and open the lid. The “Interrupted” light should go out
and the yellow “Lid Open” light should be illuminated.
4.
Remove all shipping material. Verify that the rotor assembly is correctly installed and secure on the
motor shaft.
5.
Remove the balance weights from the rotor assembly and set aside.
6.
Commence a test run by closing the lid firmly; verify that only the POWER light is illuminated; press
START, PRP or BMAC. The centrifuge should start and the automated cycle should be complete in
approximately 14 minutes. At the end of the cycle, the “DONE” indicator light should be illuminated
and audible tone sounds, indicating the rotor has come to a complete stop and it is safe to open
the centrifuge cover. The centrifuge is now ready to process blood, marrow or adipose tissue.
NOTE: Follow established protocols and procedures prior to placing the SmartPReP 2 System into service.
NOTE: If additional protective earthing (grounding) is required use the green/yellow terminal on the back
of the instrument.
CAUTION: Do not force processing disposable into rotor trunnion (swinging buckets). The processing
disposable(s) should fit snuggly but should not require excessive force to install. If resistance is
experienced, check for obstructions in the rotor and/or debris on the process disposable or
proper orientation of the process disposable to the swinging bucket.
INSTRUCTIONS FOR USE
Once the hardware is set up and has successfully completed a test run the unit is ready to operate. Any
one of the following five scenarios will automatically balance the unit if the Process Disposables (PD) are
filled properly. (Refer to the kit’s Instructions for Use for detailed filling instructions). The standard
centrifuge is equipped with a two place rotor. An optional four place rotor is available through special
order.
CAUTION
:
Do not process a combination of 20, 30 and 60mL APC or BMAC PD’s in the same cycle as they
will not balance.
CAUTION
:
Do not process APC or BMAC PD’s with the AdiPrep PD in the same cycle as they will not
balance.
CAUTION: Always have another PD or BW of equal weight in the trunnion opposing the PD to ensure
balance.