Page
16
of
26
4531200 / 1052
DECONTAMINATION PROCEDURE
AFTER BLOOD SPILLS or AS NEEDED
For disinfecting/decontamination, spray the unit with a 10% bleach/water solution and wipe with a damp
cloth that has been wet with the 10% bleach/water solution. Allow unit to dry before returning to service.
A dry cloth or paper towels may be used to dry the unit.
CAUTION: Ensure line cord is disconnected from power source prior to cleaning. Do not immerse any part
of the SmartPReP 2 System in liquid.
Before using any cleaning or decontamination methods except those recommended by the manufacturer,
users should check with the manufacturer that the proposed method will not damage the equipment.
GENERAL INSPECTION
EVERY USE
Visually check the condition of the SmartPReP 2 System. Remove from service any unit that shows signs of
physical damage or one in which the PD does not easily install. Verify that device has all four rubber feet
secured.
ROTOR ASSEMBLY MAINTENANCE
Monthly or AS NEEDED BASIS
Check movement of trunnions; there should be no binding or obstructions. Confirm the rotor nut is tight.
Hand tighten the nut, do not use a tool. Clean if necessary.
TEST RUN
Monthly or AS NEEDED BASIS
Perform verification of centrifuge process from start to finish. Commence a test run by closing the lid
firmly; verify that only the POWER light is illuminated; press START, PRP or BMAC. Confirm the system
performs a first spin, coast down and stop, second spin, coast down and stop. Process is complete in
approx. 14 minutes.
SERVICE
Harvest Technologies, Corp., or its authorized agents must perform all service. Contact Harvest
Technologies at 508-732-7500 (Toll Free 1-877-8HARVEST in continental U.S.) or your local distributor or
sales representative.
USEFUL LIFE - WARRANTY SERVICE
The SmartPReP2 centrifuge has been validated to perform as intended for up to five (5) years from the
date of purchase. Units under warranty that are received for repair, that have not been obviously abused
or impact damaged, will be promptly repaired and returned at no charge. See the Limited Warranty
section of this manual. A no-charge purchase order is requested for tracking purposes.
NON-WARRANTY WORK
Units received that have suffered obvious abuse or impact damage and units no longer under warranty will
be promptly inspected and a verbal estimate of repair cost will be provided. A purchase order is required
from the hospital that is consistent with the verbal estimate. A written estimate for repair will be provided
upon request.