Page
5
of
26
4531200 / 1052
DESCRIPTION (SMP2 with serial numbers 0101-5999)
Hardware
The SmartPReP 2 System (SN 0101-5999) is similar to a general purpose
swinging bucket centrifuge with the following exceptions:
The bucket insert accepts only the Harvest processing disposable
(PD)
The bucket orientation (horizontal or vertical) is automatically
controlled in order to perform decanting
The automated dual-spin centrifugation process is fixed and cannot
be modified by the user, thus ensuring reproducible results
The control panel is simple and intuitive, providing only three
controls: LID, START and STOP
The (LCD) liquid crystal display indicates the time remaining until the process is complete.
Control Panel
POWER
Illuminates when the unit is plugged in.
INTERRUPTED
Upon initial power up the INTERRUPTED LED will illuminate, this is normal operation.
It also illuminates when the control system senses an electronic failure. If this
condition is detected, the centrifuge is brought to a stop and cannot be restarted until
the fault condition is corrected. The INTERRUPTED LED is also illuminated whenever
the STOP keypad is pressed during a process cycle.
LID OPEN
Illuminates when LID is open or improperly closed. When this LED is illuminated, the
centrifuge is prevented from operating.
IMBALANCE
Illuminates when the control system senses an imbalance in the rotor. If this condition
occurs, the centrifuge is brought to a stop and cannot be restarted until the imbalance
condition is corrected.
TIMING
Displays the time remaining in the process cycle.