TERUMO MEDICAL PRODUCTS
Troubleshooting
A
“LOW” INDICATION
“Low” indication displays when pressurization
is not enough. It is recommended to pressurize
30–40 mmHg higher than expected systolic.
If you see the LOW sign, keep pumping to
higher number.
Make sure the tubes are
connected properly using the site window.
B
DIFFICULT PRESSURIZATION
Elemano may have a clogged air filter in
inflation bulb or there may be deterioration in
the bulb. In case of a clogged air filter, it
needs to be cleaned (refer to the Instructions
Manual, page E-18). If the inflation bulb has
deteriorated, replace it with a new bulb.
Err 1, Err 2,
Err 5
Err 3
Err 4
Errors
Symptoms
Measures
Contact Terumo technical service
Check the connection between the main
body and the cuff/inflation bulb
Reapply the cuff around the patient’s arm
Check the connection between the main
body and the cuff
Measure again while keeping the patient at
rest with the arm rested on a firm surface
Do not shake or vibrate the main body during
measurement
Use Auscultation or palpation for
measurement
Reapply the cuff around the patient’s arm
Measure again while keeping the patient at
rest with the arm rested on a firm surface
Do not shake or vibrate the main body during
measurement
Out of order
The cuff connector, the air tube
and/or the inflation bulb are
incorrectly connected to the
main body
The cuff is not applied properly
The air tube is not connected
The patient moved during
measurement
The main body was shaken or
vibrated during measurement
Abnormal blood pressure/pulse
was detected
The cuff is not applied properly
The patient moved during
measurement
The main body was shaken or
vibrated during measurement
D
Error indications
Arm circumference sizing:
Size
Centimeters Inches
SS
13-20 cm
5.1”-7.9”
S
17-26 cm
6.7”-10.2”
M
24-32 cm
9.4”-12.6”
L
32-42 cm
12.6”-16.5”
Cuff Sizes
Please refer to the Instruction Manual or visit our website for detailed troubleshooting information.
©2008 Elemano is a trademark of Terumo Corporation. TERUMO is a registered trademark of Terumo Corporation.
C
FLASHING PANEL
The Elemano display panel flashes to
warn abnormal conditions such as irregular
beats or weak pulse. If this happens, perform
the measurement in the Auscultation mode to
verify the condition.
Technical Service Hotline:
866-301-8723
www.elemano.com