Table 4
Guidance and manufacturer’s declaration – electromagnetic immunity – for
ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration – electromagnetic immunity
The LS802-B is intended for use in the electromagnetic environment specifi ed below.
The customer of the user of the LS802-B should assure that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment - guidance
Conducted RF
IEC 61000-4-6
3 Vrms
150 kHz to
80 MHz
3 Vrms
Portable and mobile RF
communications equipment should
be used no closer to any part of the
LS802-B, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
d = 1.167
d = 1.167
80 MHz to 800 MHz
d = 2.333
800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacture and d is the
recommended separation distance
in metres (m).
Field strengths from fi xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should
be less than the compliance level in
each frequency range.
b
Interference may occur in the vicinity
of equipment marked with the
following symbol:
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to
2.5 GHz
3 Vrms
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation
is affected by absorption and refl ection from structures, objects and people
a
Field strengths from fi xed transmitters, such as base stations for radio (cellular / cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast cannot be predicted theoretically with accuracy. To assess the
electromagnetic environment due to fi xed RF transmitters, an electromagnetic site survey should be considered. If
the measured fi eld strength in the location in which the LS802-B is used exceeds the applicable RF compliance level
above, the LS802-B should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as re-orienting or relocating the LS802-B.
b
Over the frequency range 150 kHz to 80 MHz, fi eld strengths should be less than 3V/m.