WARNING: No modification of this equipment is allowed.
Power supply
3.7V 1000mAH Built-in rechargeable li-polymer
battery, 6V
1A AC Adaptor
Display mode
Blue LCD with White Backlight
V.A.= 86.1mm(L) x24mm(W)
Measurement mode
Oscillographic testing mode
Measurement range
Rated cuff pressure:
0kPa-40kPa(0mmHg-300mmHg)
Measurement pressure: 5.3kPa-30.7kPa
(40mmHg-230mmHg)
pulse value:(40-199)beat/minute
Accuracy
Pressure:
5°C-40°C within±0.4kPa(3mmHg)
pulse value:±5%
Normal working condition
Temperature:5°C to 40°C
Relative humidity: ≤85%RH
Atmospheric pressure: 86kPa to 106kPa
Storage & transportation condition
Temperature:-20°C to 60°C
Relative humidity: 10%RH to 93%RH
Atmospheric pressure: 50kPa to 106kPa
Measurement perimeter
of the upper arm
About 22cm-32cm
Net Weight
Approx.284 g
External dimensions
Approx.130.9mm×73mm×29.4mm
Attachment
AC Adaptor,user manual
Mode of operation
Continuous operation
Degree of protection
Type BF applied part
Protection against
ingress of water
IP22, It means the device could protected
against solid foreign objects of 12.5 mm and
greater, and against vertically falling water
drops when ENCLOSURE tilted up to 15°
Software version
V01
Device classification
Battery Powered Mode:
Internally Powered ME Equipment
AC Adaptor charged Mode:
Class II ME Equipment
SPECIFICATIONS
38