Bionova® MiniBio
Page 5
Power supply specifications
Environment operating conditions
Input parameters
Voltage range
Frequency
Current
Output parameters
Voltage
Current
Operating
conditions
(100-240)
50/60
0,5
Values
12
2
Units
Volts
Hertz
Ampers
Units
DC Volts
Ampers
Terragene® recommends using a UPS instead of stabili-
zers, since the former has two functions: it stabilizes and
provides a continual power source during energy cutoff.
Environmental conds.
Altitude
Operation temperature
Relative Humidity
Installation/overvoltage
Degree of contamination
Storage temperature
Voltage
Functional
conds.
3500 (máx.)
10-30
30-80
category II
2
0-50
12
Units
Meters
ºC
%
ºC
DC Volts
Safety information
-Make sure that the incubator is connected to a
properly rate power cord.
To reduce the risk of using incompletely steri-
lized loads:
-Make sure to use each BI with its incubation and
reading program:
BT110
(green top) with 4 hs. 37 ºC program
BT220
(brown top) with 3hs. at 60 ºC program
BT221
(blue top) with 1 h. at 60 ºC program
BT222
(violet top) with 1 h. at 60 ºC program
BT223
(light blue top) with 30 min. at 60 ºC program
BT224
(aquamarine top) with 20 min at 60ºC
Ultra
program.
BT102
(red top) with 2 hs. at 60 °C program
BT95
(white top) with 2 hs. at 60 °C program
BT96
(grey top) with 30 min at 60ºC program.
-Check that culture medium completely wets the
spore carrier.
-Do not remove the BI until the equipment reports
the final readout result, check the impression of
the ticket.
To avoid the risk of injury, because of glass
fragments generated after the glass ampoule
inside the BI tube is crushed:
-Cool the BI during the indicated time before crus-
hing the ampoule.
-Do not handle the BI excessively since this might
cause the glass ampoule to burst.
-Wear safety glasses and gloves when removing
the BI from the sterilizer.
-Wear safety protective glasses and gloves to crush
the ampoule and to press the cap of BI.
To avoid a potentially hazardous situation:
avoid contact with the hot surface of the metal
incubator block.
To avoid BIs from absorbing fluorescent resi-
dues:
avoid BIs direct contact with chemical indi-
cators or tapes.
Do not use this product in a manner not specified
by Terragene® S.A., otherwise the protection provi-
ded by the product might be affected.
Do not replace the power supply cord for other not
provided within the product.
Only personnel authorized by Terragene® S.A can
access the inside of the incubator and its compo-
nents. Parts or components inside the machine
should not be manipulated by the user.
Serial/Batch Number
For easy identification, each Bionova® MiniBio incubator has
a unique serial number and a batch number printed on a
label found at the bottom of the incubator and on its packa-
ging. Please, register your serial number and batch number in
this manual for future reference:
Serial number:
Batch number:
Summary of Contents for BIONOVA MiniBio
Page 1: ...Rev 5 September 2018 MiniBio Innovation in technologies for Infection Control ...
Page 2: ......
Page 14: ......
Page 15: ...Rev 5 Septiembre 2018 MiniBio Innovación en tecnologías para el Control de Infecciones ...
Page 27: ......
Page 28: ......
Page 29: ...Rev 5 Setembro 2018 MiniBio Inovação em tecnologias para Controle de Infecção ...
Page 41: ......
Page 45: ......
Page 46: ......
Page 47: ......
Page 48: ......