12
Diagnostic messages on PC
Diagnostic information on LCD
Comments
Byte
Bit Nr.
Nr.
------Byte #0 - sys_diag-----
7654 3210
‘E01 Main 24 V fail’
Diag
0
0000 0010
powering voltage too low
‘E03 High internal temperature’
Diag
0
0000 1000
high internal temperature,
ensure better cooling conditions
‘E04 Temperature sensor fail’
Diag
0
0001 0000
transmitter is damaged, contact service
‘
E05 EEPROM: value check error’
Diag
0
0010 0000
reset the transmitter settings, if the
‘E06 EEPROM: check sum error’
Diag
0
0100 0000
message appears again, contact service
‘E07 EEPROM: initialized
Diag
0 1000 0000
with def. values’
————Byte #1—————————
‘E10 12V out of range’
Diag 1 0000 0001
‘E11 5V out of range’
Diag 1 0000 0010
‘E12 -5V out of range’
Diag
1
0000 0100
transmitter is damaged, contact service
‘E13 3.3V out of range’
Diag 1 0000 1000
‘E14 3.3V overloaded’
Diag 1 0001 0000
‘E15 12V overloaded’
Diag 1 0010 0000
————Byte #2—————————
‘E20 TEC unlocked’
Diag 2 0000 0001
‘E21 TEC overvoltage’
Diag 2 0000 0010
‘E22 TEC overcurrent’
Diag 2 0000 0100
‘E23 LD failed’
Diag
2
0000 1000
transmitter is damaged, contact service
‘E24 LD degrade’
Diag 2 0001 0000
‘E25 LD shut down failed’
Diag 2 0010 0000
‘E26 LD overcurrent’
Diag 2 0100 0000
‘E27 Critical temperature’
Diag
2
1000 0000
ensure better cooling conditions
————Byte #3—————————
‘E30 Low RF level’
Diag
3
0000 0001
increase RF level or set the attenuator
“Attn” level properly
‘E31 High RF level’
Diag
3
0000 0010
decrease RF level or set the attenuator
“Attn” level properly
‘E32 Hazardous RF level’
Diag
3
0000 0100
decrease RF level and reset internal
attenuator
“Attn”
‘E33 AGC out of range’
Diag
3
0000 1000
input RF level cannot be compensated
by internal attenuator
Table 1
The diagnostic information is upgrated after power up or transmitter reset.
Note:
Bit=1 Error, Bit=0 __ Ok.