16
10. PROGRAMMES
10.1. PROGRAMME SETTINGS
Programme Frequency
(Hz)
Pulse Width
(µs)
Ramp
(sec)
Work
(sec)
Rest
(sec)
Default
duration
(min)
1
STRE 1
50
300
1
5
10
20
2
STRE 2
35
250
2
3
6
20
3
URGE
10
200
1
5
10
20
4
MIXED
10/50
200/300
30
5
URGE 2
10
200
Constant
Continuous
6
SENS
3/10/20/30/40
250/200
25
7
PFW A
See the table below
8
PFW B
See the table below
9
ENDUR
20
250
2
5
10
20
10 PAIN
4
200
Constant
20
11 TIBN 1
10
200
Constant
20
12 TIBN 2
20
200
Constant
20
13 CSTM 1
4-90
50-400
1-10
1-20
0-30
5-60
14 CSTM 2
4-90
50-400
1-10
1-20
0-30
5-60
15 CSTM 3
4-90
50-400
1-10
1-20
0-30
5-60
.
PFW
Frequency
(Hz)
Pulse Width (µs)
Work(sec)
Rest (sec)
Default
duration (min)
A
20
250 µs fixed
4
6
2
20
250 µs stepped
4
6
5
3
250 µs fixed
4
4
5
10
500 µs fixed
4
4
15
20
250 µs stepped
4
6
10
35
250 µs stepped
4
8
5
45
250 µs stepped
4
8
5
10
500 µs fixed
4
4
5
Total
52 min
B
20
250 µs fixed
6
8
2
20
250 µs stepped
6
8
5
3
250 µs fixed
6
6
5
10
500 µs fixed
6
6
15
20
250 µs stepped
6
8
10
35
250 µs stepped
6
12
5
45
250 µs stepped
6
12
5
10
500 µs fixed
6
6
5
Total
1. min
In PFWA the Pulse Width increases from 175µs to 250 µs in 4 secs - 25 µs per sec. This stepping up
occurs during Output ON time.
In PFWB the Pulse Width increases from 170 µs to 250 µs in 8 secs - 10 µs per sec.