BATTERY AND FUSE REPLACEMENT
•
If icon appears on LCD, please replace the battery as follows:
•
Disconnect test probes from circuits being measured, turn range selector to OFF
position.
•
Remove the screw from the battery cover, and separate the battery cover from the
rear casing.
•
Replace the 9V
battery with a new one
(6F22) noting the correct polarity.
•
Refit the battery cover and tighten
the screw.
Battery
Replacing the Fuses
Warning:
To avoid electrical shock, personal
injury or damage to the Meter, use specified
fuses ONLY in accordance with the following
procedure.
•
Turn the rotary switch to OFF and remove
all connections from the terminals.
•
Remove the 5 screws from the case
bottom.
•
Remove the fuse by gently prying one
end loose, then lift out the fuse from its
holder.
•
Install ONLY replacement fuses with the
identical type and specification as follows
and make sure the fuse is fixed firmly in
the bracket.
•
Fuse 1: 0.5A,1000V, fast type 6.3x32mm
Fuse 2: 10A,1000V, fast type 10.3x38mm
•
Refit the case bottom and case top, and install the 5 screws.
Note:
Replacement of a fuse is seldom required, a blown fuse is nearly always the
result of improper operation.
Fuse