12
Storing and Clearing Readings
•
To store readings, proceed as follows:
1.
Press STORE once, STO and “no.xxxx” displays to confirm the operation and the
left secondary display shows the current measurement reading. Press to toggle
between clearing the stored readings and start from the first readings or start from
the last stored reading. Right secondary display shows the original number of
records.
2. Press STORE a second time and STO displays. The left secondary display shows
the storing time interval in seconds, it is preset to zero.
3. To change the interval in seconds by pr or - button. The interval can be as
high as 255 seconds or as low as 0 second.
4. Press and hold STORE to access the quick setting.
5. Press STORE the third time, STO and no displays. The left secondary display
shows the index number increase by one. The right secondary display shows the
value of the corresponding index number, the primary display shows the current
measurement reading.
•
If there is no interval time set to store the reading, each press of STORE
increments the reading by one index number.
•
The maximum number of stored reading is 100. When the stored readings memory
is full, the meter will start to revert to the first stored reading and overwrite.
•
To exit, press EXIT.
•
Automatic power off feature is disabled while the meter is in this mode.
Recalling Stored Readings
•
Use the following procedure to recall the stored reading:
•
Press RECALL to recall the stored value and RCL displays to confirm the
operation.
•
The left secondary display shows the index number “no.xxxx”.
•
The primary display shows the corresponding recalled data.
•
The right secondary display shows the total number of the stored data.
•
Press button to enable the SEND feature to export the data to the computer via
USB. The software shows the data storing time and also the data value. After the
data transferring is completed, the SEND feature will be disabled automatically.
•
Press + or - button to view additional stored reading.
•
Press and hold RECALL to access quick recalling.
•
Press EXIT to exit recalling.
Power measurement
•
Set the rotary switch to W.
•
Insert the power adaptor to the corresponding input terminals, and plug the power
adaptor to the outlet.
•
Insert the object to be measured into the outlet of the power adaptor.
•
The measured value shows on the display. The value includes: apparent power
value, power value and power factor value.
•
When power measurement has been completed, first switch off the power, then
disconnect the connection between the adaptor and the outlet.
Note:
The current of the object being measured must >10A.
≤
5A continuous
measurement is allowed. When measuring 5A~10A, only
≤
10 seconds continuous
measurement is allowed and the interval between each measurement must be greater
than 15 minutes.