background image

 

 

TEN 

 
 

 

Test Equipment Nederland B.V. (C)      Version 3.0  February 2008 

Pagina 10

 

  5  Messages 

 
 
 
During operation, the EDA 500 can display different messages. The following messages can occur: 
 

EDA2   

No communication

 

between the IINOVA 500 and EDA2 smoke head. Check the communication  

OFF-LINE

 

cable connection. Check that the EDA2 is switched on. 

 

EDA2   

The smoke cell temperature is below 80  ºC. The smoke cell is heating up. 

HEATING

 

 

 
ERROR 

This message can occur during a zero calibration. Reason for the error can be: 

 

 

a) Receiver needs to be cleaned, see Chapter 12. 

 

 

b) The halogen light source is defect, see Chapter 12. 

 

 

c) The smoke head is located in a location with to much direct sun light. 

 

 

    Move the smoke head to another location. 

 

 

  6  Maintenance 

 
 

Cleaning 

Clean the outside of the INNOVA 500

 

and smoke head with a soft towel and eventually with a non aggressive liquid 

like Glassex. Before cleaning the unit always switch off the INNOVA 500

 

and remove the mains plug. Let the smoke 

head cool down till ambient temperature. 
 

Replacing the printer paper 

Open the printer by pulling the lit at the cover forwards. Remove the old roll and place the new roll in the printer. Let 
a piece of paper of approximately 10 cm hang out of the printer. Press with both hands at  the printer cover to close 
the printer. Press a the paper feed key to check the paper transport.  
 

Replacing the fuses 

Before you change the fuses always switch off the INNOVA 500, remove the power lead and let the smoke head 
cool down until ambient temperature. Open the little lit of the main switch at the rear of the EDA, no. 12 of figure  2. 
Take out the fuses and place the new fuses in the holders of the lit. Use 230 V 2AT fuses. Close the lit. 
 
In the smoke head the fuse holder is located at the bottom of the EDA 2, no. 5 of figure 3. Press and open the 
holder by turning it counter clockwise. Take of the holder with the fuse. Replace the fuses. Use 230 V 2AT fuses. 
Close the fuse holder by pressing it and turning it clockwise. If you have to change the fuses frequently, please 
contact our service department! 
 

Calibration 

The EDA 500

 

must be calibrated once every 12 month. The calibration can only be performed by authorized 

technicians from your supplier. We advise you to take a service contract including the annual calibration.  
 

Cleaning the optical components of the smoke head  

If the receiver gets polluted, the receiver voltage will go down and the smoke meter will show ‘ERROR’ during a 
zero calibration at the displays. Switch off the INNOVA 500

 

and let the smoke head cool down until ambient 

temperature. Loosen the right mounting ring of the smoke chamber and take out the cone. Now take the receiver 
support out of the cone. Clean the front surface of the receiver with a clean and soft cotton towel. Don’t touch this 
surface, the lens, with your fingers. Put the support back to its position and place cone into the smoke chamber. 
Tighten the mounting ring. Switch on the smoke meter and perform a zero calibration. Check that the smoke meter 
is not showing the error message again. 
 

Summary of Contents for EDA 2

Page 1: ...ftware are the intellectual property of Test Equipment Nederland B V Each way of duplication of this manual or software is not allowed Test Equipment Nederland B V is not responsible for any direct or indirect damage arising from the use of the apparatus or manual The EDA 2 is tested and approved according the EMC directive 89 336 EC ...

Page 2: ...TEN Test Equipment Nederland B V C Version 3 0 February 2008 Pagina 2 Contents 1 Introduction 3 2 Preparations 4 3 Smoke test 6 4 Official test 8 5 Messages 10 6 Maintenance 10 7 Specifications 12 ...

Page 3: ...completely made of aluminum and uses the so called unique natural flow system that guarantees a minimum of maintenance In this user manual we describe the combination of the EDA2 smoke head and the INNOVA500 The INNOVA 500 has digital LED displays and a thermal printer to provide you all necessary information during a emission test Before using the EDA 2 it s absolutely necessary to read this user...

Page 4: ...RPM is recorded correctly If not repeat the procedure mentioned above at a different location of the injection pipe Figure 2 Placing the piëzo clamp Remark the piëzo clamp can not be used for all types of diesel vehicles for example common rail diesels Here for you need a special RPM adapter If you use an adapter always follow the instructions given by the manufacturer Oil temperature sensor Take ...

Page 5: ...ameter equal or larger than 70 mm Place the probe a far as possible in the exhaust Take care the gas flow in probe is correct Avoid sharps curves or bending Figure 3 Placing the sample probe The temperature of the sample probe can rise up to 200 ºC during a smoke test Always wear protection gloves while you are touching the exhaust or sample probe During a zero calibration of the smoke meter the s...

Page 6: ...displayed If the opacity and max opacity display shows the message you need to perform a zero calibration Press the zero calibration key to activate a zero calibration Figure 5 Zero calibration of the smoke head Important We advise you to perform a zero calibration regularly Take care that during a zero calibration the sample is removed from the exhaust of the vehicle Printing If you wish to print...

Page 7: ...during a test always use MODE Select mode A not filtered or mode B filtered Normally during a test always use mode B PROBE Select sample probe 1 or 2 Probe 1 Ø 10 mm exhaust diameter 70 mm passenger cars Probe 2 Ø 27 mm exhaust diameter 70 mm large commercial vehicles DIVISION Here you can set the division ratio for the RPM for example 1 2 3 4 etc DISPLAY Select what you want to display at the dis...

Page 8: ...l a temperature of 60 ºC is reached As soon as the engine temperature reaches 60 ºC the EDA automatically continues If the temperature of 60 ºC is not reached or you just want to continue press the Enter key Now the INNOVA 500 will proceed with the next program part After measuring the temperature the operator must be prepared to perform the free accelerations Between each acceleration the operato...

Page 9: ...Nederland B V C Version 3 0 February 2008 Pagina 9 Press the print key to get a full report of the official test with all recorded values Press the escape key to close the official test and return to the normal smoke test ...

Page 10: ...A 500 remove the power lead and let the smoke head cool down until ambient temperature Open the little lit of the main switch at the rear of the EDA no 12 of figure 2 Take out the fuses and place the new fuses in the holders of the lit Use 230 V 2AT fuses Close the lit In the smoke head the fuse holder is located at the bottom of the EDA 2 no 5 of figure 3 Press and open the holder by turning it c...

Page 11: ... of the new light source at the bottom but leaf the plastic around the bulb Carefully press the new light source with its two pins at the bottom into the holder of the support Remove the plastic cover Never touch the light source with you fingers Place the support into the cone and put the cone into the smoke chamber Tighten the mounting ring Switch on the smoke meter and perform a zero calibratio...

Page 12: ...30 mm Material Aluminum Diameter 55 mm Light source Halogen 6V 10W 3200 ºKelvin Receiver Siemens BPW21 with V filter Sample probe 10 mm 1 meter length Optional 16 mm 2 5 meter length 27 mm 1 meter length trucks 27 mm 3 5 meter length trucks Ranges Opacity 0 99 9 0 9 99 m 1 Oil temperature 0 150 ºC Cell temperature 0 150 ºC RPM 300 9990 1 min Resolution Opacity 0 1 0 01 m 1 Oil temperature 1 ºC Cel...

Reviews: