background image

 

 

 

USER MANUAL 
UM-PDUB-181030 REV.A

Page  20  /  23 

FAQS 

Why sometimes remaining battery of Tempod logger becomes empty very soon?

 

 

Sometimes incorrect installation of battery may make logger not able to correctly recognize the remaining battery, you could try steps below to 

make logger correctly recognize the remaining battery. 

 

1. 

Remove battery from logger

 

2. 

Press the “View/Mark” button for several times until LCD changes to blank

 

3.

 

Install battery back to the logger

 

4.

 

Re-program logger if LCD shows error code E009

 

 

 

Summary of Contents for Tempod 100X

Page 1: ...USER MANUAL UM PDUB 181030 REV A Page 1 23 USER MANUAL TempSen Tempod USB Data Loggers UM PDUB 181030 REV A TEMPSEN ELECTRONICS COMPANY www tempsen com...

Page 2: ...iew 3 Configure Tempod 6 Start Recording 9 LCD Screens 11 Download Data 13 Stop Recording 13 Time Mark 14 Battery 15 Data Results Explanations 16 Additional Information 18 TempCentre Data Management S...

Page 3: ...pod 30 Tempod 50X Tempod 100X Tempod 200X Sensor Temperature Internal NTC sensor Temperature External NTC sensor probe 1m Temperature External RTD sensor 1m Temperature External RTD sensor 1m Memory C...

Page 4: ...Empty no alarm setting 2 Triggered Alarm Zones Hi High Lo Low A1 A2 3 Remaining Battery Indicator 4 AccessKey Password protected 5 Statistics 6 Unit C F d day h hour m minute s second 7 Time Mark 8 St...

Page 5: ...protection against harmful interference in a residential installation This product generates u ses and can radiate radio frequency energy and if not installed and used in accordance with the instructi...

Page 6: ...Disk configuration is done Configuration Options A unique 10 digits alphanumeric number automatically created for identifying each monitoring task LCD displays USB when logger connected to PC LCD disp...

Page 7: ...ed temperature conditions Device automatically starts recording on the specific date time preset no need to press Start Stop button after configuration to device No alarm setting High Low alarm 2 alar...

Page 8: ...ressing Start Stop button recording stops when logging cycle completes however you are still allowed to stop recording through TempCentre or TempSen Go APP appears on LCD after Stop Button was disable...

Page 9: ...er LCD Screens after Configured Manual Start Start Delay LCD displays Stby after configured Logger was set to start option of start deley Manual Start Temperature Conditions LCD displays Stby and Temp...

Page 10: ...g begins after start delay time passed Manual Start with Temperature Conditions Logger begins to measure temperature and returns readings to LCD after logger was manually activated recording begins wh...

Page 11: ...Current Reading 15 3 C Status Recording No Alarm triggered Current Reading 15 3 C AccessKey enabled Status Recording Alarm triggered at alarm zones Hi Lo Current Reading 26 4 C Status Recording Alarme...

Page 12: ...UM PDUB 181030 REV A Page 12 23 Statistics Press View Mark button on data logger to view recording statistics Max Maximum value Min Minimum value Avg Average value Max Value 37 5 C Time Marks 3 of 9 m...

Page 13: ...nagement platform www tempsencloud com for further analysis Logger would be ready for USB communication with TempCentre software only after PDF report is completely created Download Data via TempCentr...

Page 14: ...tion is effective and please keep unit recording stops Recording Stops StoP displays after recording stopped Press View Mark button on data logger to place a time stamped mark up to 9 time marks avail...

Page 15: ...c CR2450 battery into battery compartment with positive side upward Put cover back to compartment in the position where triangle on cover points to the triangle on logger then spin cover in the direct...

Page 16: ...ording duration time which is calculated from the first point to the last point MKT Mean Kinetic Temperature is a simplified way of expressing the overall effect of temperature fluctuations during sto...

Page 17: ...ts Number of violation event for each alarm zone shows in Events column of alarm statistics in data report The total time that readings outside of each alarm limit which equals to number of points out...

Page 18: ...store to normal status To avoid damaging your data logger please do not disconnect it from USB port while it is communicating with the computer Do not use any aggressive cleaning agents or solvents to...

Page 19: ...mation required prior to downloading then an email containing download link would be sent to the mail address you left in couple of minutes If you want to download the software from www tempsen com ag...

Page 20: ...of battery may make logger not able to correctly recognize the remaining battery you could try steps below to make logger correctly recognize the remaining battery 1 Remove battery from logger 2 Pres...

Page 21: ...0004 Error 004 Device Rest E006 Error 006 Start time error E009 Error 009 Start failure resulted from low battery E101 Error 101 CFG file checksum error E102 Error 102 CFG file error incorrect device...

Page 22: ...rademark trade name or logo affixed to it The Limited Warranty does not apply to any non TempSen hardware product or any software even if packaged or sold with the TempSen hardware Manufacturers suppl...

Page 23: ...REV A Page 23 23 REVISION HISTORY Revision Date Details A 2018 10 30 First version TempSen Electronics Company T 86 21 6768 5210 F 86 21 6768 5232 E info tempsen com www tempsen com Technical Support...

Reviews: